background image

10

Chin Rest 

Wire 

Base Lamp

Wire

 Figure- 8 Connections

Figure- 6 Illumination & Moving Arm Assembly

Figure- 5 Install Base

Figure. 7 Assembling Power Supply on the

Power Supply Bracket

Move 
Handles 
up or 
down

Allen Cap 
Screw

Figure- 9 Adjusting Patient

Handle Height 

Figure- 7 a. Earthing wire attaching 

screw

 SL 20i 

SETUP (continued)

8126700-795 

 SL 20i 

Issue 1.0

Printed on June,2019

UNPACKING AND INSTALLATION (continued) 

7.  Install the Slit Lamp Base Assembly onto the track
    of the Table Top and slide the Guide Rail Covers
       around the tracks. Refer to Figure-5. Lock the base
       at any position to prevent any undesired movement 
      by using Base Locking Knob Shown as (A) in Fig. 5.            

8. Assemble Illumination & moving arm assembly
    onto the base using Allen Screw M6 as shown in
      Figure.6.      
 
9. Assemble Power Supply to the table top just by
     pushing into the hole shown as (C), refer Figure-7.

10.Connect the power supply earthing wire with table
    top at point “D”and “E”using 3mm Allen Wrench. 
      Refer Figure-7&7a.  

11.Attach the Base Lamp Wire Connector to the back
      of the Power Supply Assembly. Refer to Figure-8.

12.Connect the Chin rest target light Connector to the
     Power Supply Assembly by plugging it in. Refer to     
      Figure-8.

13. Using the 3mm Hex Wrench, Assemble the Patient
   Handles to the Platform Assembly as shown in 
      Figure.9.       

C

D

E

Slide

A

Summary of Contents for SL 20i

Page 1: ...Manual Part No 8126700 795 Issue 1 0 Printed on June 2019 To ensure proper use of this instrument as well as to avoid injury while operating Instrument understanding this manual completely before use...

Page 2: ...Reserves the right to make change in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 13485 Certified LABOMED products are d...

Page 3: ...ION OF INPUT POWER 11 INSTRUCTIONS FOR USE 12 OPERATION 12 SLIT LENGTH 13 FILTER DIAL 13 SLIT ROTATION 13 CLEANING MAINTENANCE 14 EXTERNAL CLEANING 14 FOREHEAD CHIN REST PREPARATION 14 CLEANING THE GL...

Page 4: ...O ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING TO AVOID RISK OF ELECTRIC SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAIN WITH PROTECTIVE EARTH OTHERWISE FAILURE OF INSTRUMENT AND...

Page 5: ...AY OCCUR SEE THE CARE AND MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO ELECTROMAGNETIC CHARGE EMC AND NEEDS TO...

Page 6: ...Power Protective Earth Connection ON OFF Manufacture Catalog Number Serial Number Waste of Electrical and Electronic Equipment Compliance to medical Device Directive 93 42 EEC Accompanying Document m...

Page 7: ...ticians and eye care technicians should operate this instrument Please retain this manual for future reference and to share with other users Additional copies can be obtained from your authorized LABO...

Page 8: ...k Knob 18 Instrument base Lock Knob 19 Guide Rail Cover 20 Geared Rollers 21 Guide Rails 22 Fore head support 23 Power Supply Assembly 24 Table Top 25 Patient Handles 26 Observation Head 27 Chin Suppo...

Page 9: ...g Item Packed as Shown in Figure 2 a Chin Rest Assembly b Microscope Head with Eyepiece Mounted c Power Supply Box d Base Assembly e Patient Resting Handle f Illumination Arm Assembly g Accessories Sl...

Page 10: ...ck the base at any position to prevent any undesired movement by using Base Locking Knob Shown as A in Fig 5 8 Assemble Illumination moving arm assembly onto the base using Allen Screw M6 as shown in...

Page 11: ...MINER OR A DANGER TO THE PATIENT WARNING POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION DEVICE PLUG 1 After the unit is in its secure location apply the correct inpu...

Page 12: ...bs adjust the projected slit so that the thinnest slit is shown on the Focusing Rod Refer to fig 12 15 NOTE The thinnest line will allow for greater accuracy 6 Remove the Focusing Rod 7 To position a...

Page 13: ...dicated in the scale of the slit lamp arm Refer to figure 17 15 Magnification is altered by using lever on the Magnichanger Refer to figure 16 SLIT LENGTH The Slit length is adjusted by rotating the S...

Page 14: ...dampened with a mild soap and water solution CHANGING THE HALOGEN BULB WARNING NEVER REMOVE A BULB THAT HAS RECENTLY BEEN IN USE AS IT WILL BE VERY HOT WAIT UNTIL IT HAS COOLED AND USE GLOVES OR A TH...

Page 15: ...y FUSE REPLACEMENT Replace the fuses in the Power Input Module with the fuses indicated in the Specifications section of this manual 1 Remove input power to the instrument 2 Press down on the top tab...

Page 16: ...bulb housing Double slit visible in microscope Microscope not focused on focusing rod before use Install focusing rod and check to ensure microscope is focused on it Bulb not installed properly Check...

Page 17: ...80 Atmospheric Pressure 80 kPa 23 6 in Hg to 106 kPa 31 3 in Hg Transportation Storage 10 Temperature 20 C 4 F to 70 C 158 F 106 kPa Relative Humidity 10 to 80 non condensing Atmospheric Pressure 50k...

Page 18: ...nnecting the equipment to a protective grounding system so that the accessi ble metal parts do not become conductive in case of failure in the basic insulation Grade of protection against a hazardous...

Page 19: ...T H S ADAPTER KIT SBS WITH VIDEO OUTPUT CCD ADAPTER 8122500 8122600 800 6134131 6134110 6134120 6134172 I PHONE 6 6134172 1 6 I PHONE 6 PLUS 6134172 1 6P 8122600 851 8122600 852 CCD CAMERA I PHONE 7 6...

Page 20: ...Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class A Harmonics IEC61000 3 2 Class A Flicker IEC61000 3 3 Complies 8126700 795 SL 20i Issue 1 0 Printed onJune 201...

Page 21: ...5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds Power Frequency 50 60Hz Magnetic Field IEC61000...

Page 22: ...ccording to the transmitter manufacturer and dis therecom mended separation distance in meters m Field strengths from fixed Rf transmitters as determined by an electromagnetic site survey should be le...

Page 23: ...or use in the electromagnetic environment in which radiated RF disturbance are controlled The customer or user of the SL 20i can help prevent Electromagnetic interference by maintaining a minimum dist...

Page 24: ...t Labomed shall not be liable for any special Incidental or consequent damages for any negligence breach of warranty strict liability or any other damages resulting from or relating to design manufact...

Page 25: ...Revision History Rev No Date of Release Change App By DCR...

Page 26: ...4538 U S A Phone 510 445 1257 Fax 510 991 9862 E mail sales laboamerica com www laboamerica com Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10 4584222 Fax 31 0 10...

Reviews: