Prima ENT/DNT
4.
Remove the inclined coupling assembly from its
packaging and follow the instructions below (see
Fig. 4.)
A. Install the coupling by sliding the shaft (1)
into the suspension arm.
B. Lock the inclined coupling with the threaded
knob (3).
C. Make sure the safety screw is installed and
tightened at position (2) so the coupling
does not detach if knob (3) is accidentally
unscrewed.
5.
Mount the light guide by completing the steps
below.
A. Switch off the illumination system.
The light guide comes pre-routed through
the swivel arm (which houses the LED
illumination system) and the suspension arm.
B. Remove the dust cap from the receptacle.
C. Insert the light guide into the receptacle in
the microscope until it clicks into position as
shown in A in Fig. 5.
D. Make sure the light guide has been routed
in such a way that the carrier system and
the surgical microscope are not obstructed
and can be moved in their entire range of
movement without stretching, kinking, or
twisting the light guide.
6.
Install the binocular head and eyepieces on the
magnichanger. Secure the binocular head with
the head locking screw as shown in A in Fig. 6.
Fig 4
Fig 5
Fig 6
3
1
2
Safety Screw
A
A