Outer Cover
8 Changing bulb for incident & Transmitted Lamp
9 Changing bulb for Transmitted illumination
CZM6 Stereo Zoom Microscope User Manual
7
1
Use any 12V-20W Halogen with reflector dia 35mm
for incident light (GE, OSRAM bulbs suggested).
2
Turn the microscope OFF. Unplug the Microscope by
removing the power cord. Verify once again that the
main power switch is in OFF position.
3
Open the outer cover of the incident light housing by
anti-clockwise rotation.
4
Using tissue paper, replace the old bulb and insert
new bulb in bulb cavity.
CAUTION:
To avoid any injury, please make sure
microscope has been off for at least 15 minutes
before removing old bulb.
•
Use any 12V - 10W Halogen Bulb for transmitted
light (GE, OSRAM bulbs suggested)
•
Repeat step 2 as mentioned above.
•
•
Repeat step 4 above.
CAUTION:
To avoid any injury, please make sure
microscope has been off fro at least 15 minutes
before removing old bulb.
Remove observation head and turn microscope
upside down in order to access bulb housing through
the base plate. Pull on locking knob to open the
lamp door.