OPERATING INSTRUCTIONS

• 

Turn ON the illumination system of the Microscope.

Place the observation specimen on the glass plate of the Analyzer.

For initial setup of the instrument, the fiducial line of the polarizer 

should coincide with 0 on scale. At this position the analyzer and 

polarizer set should show the brightest field. 90° left or right of this 

position with yield darkest field.

Adjust the required angle of incident polarized light using the scale 

provided on the polarizer while viewing the image through the 

eyepieces.

Rotate the polarizer slowly to achieve the required position.      

CARE & MAINTENANCE

?

?

?

Loose particles can be removed from the surface of the Analyzer and Polarizer by gently wiping

them with lens tissue dampened with Acetone or Methanol.

It is better to avoid leaving finger prints on the surface of Analyzer and Polarizer.

When not in use for long time, it is recommended to store the polarizing kit in its original packaging.   