BA_057_en_2022-03_11.00  Electronic angle-of-rotation sensor 81.4 

Page 4/5 

4.1.1 

Zero-point correction 

Zero point deviations can occur due to the 
mounting, the installation position or due to 
the type and duration of use. Measuring de-
vices with micro adjustment pointer can be 
corrected  in  a  depressurised  state  by 
means of the adjusting screw on the pointer 
hub (see figure 1).  

You can find further information about zero-
adjustment  of  pressure  gauge  with  micro 
adjustment pointer in the document TA_029 
on www.labom.com.  

 

Figure 1: Zero-point correction 

  

4.1.2 

Adjusting the angle-of-rotation sensor 

For  adjustable  devices  the  adjustment  re-
quires to apply the lower limit pressure re-
spectively the upper limit pressure to the de-
vice. Bring the enclosed or another magnet 
in  the  alignment  position  (see  figure  2)  to 
adjust accordingly. After approx. 2 seconds 
a light signal (1) in the scale acknowledges 
the  adjustment  of  the  measurement  range 
begin  and/or  the  adjusted  range.  The  ad-
justment is possible within a tolerance of ap-
prox. ± 5 % F.S. Any settings that differ from 
this will not be accepted. 

 

  

Figure 2: Alignment position 

  

4.2  Maintenance / Service 

When properly installed in accordance with applicable specifications, this device is mainte-
nance-free. However, we recommend an annual recalibration of the device. 

  

In the event of any damage or defect the customer cannot replace or repair any components 
or assemblies. 

  

 

 