background image

Operating Instruction BA_029_2019-08_10.00  Pressure transmitter PASCAL CV

Page 4/6

 

BTA-No. 029  1J2  Page 4

Operation of pressure transmitter PASCAL CV with Switching Module 

 The standard factory setting for the basic module and the switching module can be found in the table on page 5. 

Internal module 

plug connection

x) 

 

x)   Various function modules can easily be added to PASCAL CV  for displaying or communicating purposes  

(see table p. 5). 

Setting lower range value (Operating menu: Zero Point)

1.  Create a stable pressure at which 4 mA is to be set. 

2.  Hold down the left button (ZERO) for about 3 seconds. 

3.  The change to the lower range value is acknowledged after you release the button by a flashing “ZERO” 

LED. 

Note:

The measuring span is always retained when the lower range value is changed if this is permitted by the 

sensor. A completed change is acknowledged by a flashing “ZERO” LED.

 

 

If the pressure transmitter is set outside the permissible measuring range limits (nominal range ±5 %, smallest 

measuring span) then the values will not be applied and the LED will not be acknowledged. 

 

Setting upper range value (Operating menu: Measuring Span)

 

1.  Create a stable pressure at which 20 mA is to be set. 

2.  Hold down the right button (SPAN) for about 3 seconds. 

3.  The change to upper range value is acknowledged after you release the button by a flashing “SPAN” 

LED. 

 

 

Note:

The lower range value is always retained when upper range value is changed. A completed change is 

acknowledged by a flashing “SPAN” LED.

 

 

If the pressure transmitter is set outside the permissible measuring range limits (nominal range ±5 %, smallest 

measuring span) then the values will not be applied and the LED will not be acknowledged. 

 

 

Channel 1

 

1.  Apply the pressure at which the switch is to be activated. 

2.  Hold down the left button “ZERO” (for switching channel 1) for about 10 seconds. The “ZERO” LED then lights up 

continuously. 

Abort: press the “ZERO” button again for about 10 seconds. Alternatively, wait for an automatic abort after about 5 

minutes. 

3.  If the pressure is stable, press the “ZERO” button for a short time. 

4.  The switching point is accepted and the “ZERO” LED stops flashing. 

 

Channel 2

 

1.  Apply the pressure at which the switch is to be activated. 

2.  Hold down the right button “SPAN” (for switching channel 2) for about 10 seconds. The “SPAN” LED then lights up 

continuously. 

Abort: press the “SPAN” button again for about 10 seconds. Alternatively, wait for an automatic abort after about 5 

minutes. 

3.  If the pressure is stable, press the “SPAN” button for a short time. 

4. 

The switching point is accepted and the “SPAN” LED stops flashing.

 

Setting the measuring range

 

BTA-No. 029 Rev. 0F3

P. 6

Module mounting 

LED „SPAN“ 

acknowledgement 

Button “SPAN” for 

setting upper range 

value and switch 

point channel 2 

LED „ZERO“ 

acknowledgement 

Button “ZERO” for 

setting lower range 

value and switch 

point channel 1 

Overload indicator 

for channel 2

Terminals 

Module mounting 

Overload indicator 

for channel 1 

Setting the switch points (Operating Menu: Switch Points)

 