Instruction Manual |
1
1.0 Introduction
The Labnet Rocker 35 EZ is a general purpose adjustable tilt rocker with digital time and speed controls. The tilt angle is easily
adjustable without tools using the knob on the right side of unit. This reliable rugged unit is designed for extended use and comes
with a 5-year warranty.
The motor is designed to generate a rocking effect, while the electronic interface controls the RPM and Time features.
2.0 Specifications
The Labnet Rocker 35 EZ housing is made of plastic and steel plate varnished with high resistant polyurethane lacquer.
Electrical
120V
230V
120V ± 10%, 50/60Hz, 0.3A
230V ± 10%, 50/60Hz, 0.15A
Motor Power
6 W
Fuse
2 x 1AT
RPM Regulation
Microcontroller, load independent, from 10 to 120 rpm
Tilt Angle
0° to 20° from horizontal
Timer
10 to 120 min., HOLD function
Load
Max. 11 lbs. (5 kg)
Dimensions (W x L x H)
Single platform
Double platform
14.2 x 14.2 x 6.7 in. (360 x 360 x 170 mm)
14.2 x 14.2 x 10.2 in. (360 x 360 x 260 mm)
Platform Dimensions (W x L)
14.2 x 11 in. (360 x 280 mm)
Weight
Single platform
Double platform
14.3 lbs. (6.5 kg)
17.2 lbs. (7.8 kg)
Operating Conditions
4°C to 65°C, up to 85% RH, non-condensing
This device is intended for use in a basic electromagnetic environment (Basic, class B), and it is in compliance with IEC 61326-1.
The power cord can be exchanged only with a cord 3 x 1.0 mm
2
or for USA/Canada 3 x 18 AWG.
The Labnet Rocker 35 EZ is designed to be safe when operated under the following conditions:
Indoor use
Altitude up to 2,000 meters
Pollution Degree 2
3.0 Safety Precautions
Do not use the rocker near water. Ensure that water does not get on the device, particularly during cleaning.
Do not rock flammable or explosive samples.
If the unit is used in a manner not specified in this manual, then the unit may become unsafe to operate and may void
the warranty.
This product must be connected to a grounded power outlet for safe functioning.
Use only the power cord supplied with the product.
The equipment will not restart if the main power source is interrupted.
The equipment will restart in cases of mechanical interruption.
Do not expose the rocker to aggressive vapors (e.g., organic solvents).
Do not insert any objects into the rocker (such as a screwdriver or metal rods).
Never operate the rocker without the platform securely attached.
Hazardous samples should be shaken only in appropriate containment vessels.
The electrical warning symbol indicates the presence of a potential hazard which could result in electrical shock.
Do not put your fingers between the rocking platform and the housing. Danger of mechanical injury.