 8.0  Troubleshooting and Repair

Problem

Probable Cause

Remedy

No display shown in the 
monitor when the power is on.

Power is not connected.

Check the power and connect it.

Fuse has burned out.

Change the fuse:

 



BYQ15412A000020: 125V, 2.5A, 5 x 20 mm

 



BYQ15411A500170: 250V, 1.5A, 5 x 20 mm

Switch is broken.

Change the switch.

Other.

Contact the supplier or manufacturer.

During Run, the housing shakes 
abnormally and makes an 
abnormal noise.

Use improper block.

Change the block (contact the supplier or manufacturer to 
make sure whether the block is suitable).

Unit is placed on uneven surface.

Place the unit on an even, stable surface.

Round feet badly worn.

Change round feet.

Other.

Contact the supplier or manufacturer.

Bad temperature uniformity of 
the block.

Improper installation of the block. Reinstall the block.
Bad cooling components.

Contact the supplier or manufacturer.

The system no longer operates 
after changing the block.

Bad grounding of the machine.

Check the power cable to ensure reliable grounding, 
remember to turn off the power when changing the block.

Serious discrepancy between 
the actual temperature and the 
displayed temperature.

Broken sensor or bad contact.

Contact the supplier or manufacturer.

The cooling of the module 
drastically slows down or the 
temperature cannot reach 
below room temperature.

Thermoelectric cooling module is 
broken.

Contact the supplier or manufacturer.

Fan is broken or does not work.

The module is neither able to 
heat nor cool.

Temperature sensor is broken.

Contact the supplier or manufacturer.

Thermoelectric cooling module is 
broken.

Vibration stops.

Drive is broken.

Contact the supplier or manufacturer.

Motor is broken.
Fuse has burned out.

Change the fuse:

 



BYQ15412A000020: 125V, 2.5A, 5 x 20 mm

 



BYQ15411A500170: 250V, 1.5A, 5 x 20 mm

User shall not open the cover of this instrument when under warranty. If housing does need to be opened to perform 
certain repairs per the above instruction, please contact the supplier or manufacturer.

Should you have a question about the operation of the Labnet Microtube Shaking Incubator or if service is required, contact Customer 
Service. Do not send in a unit for service without first calling to obtain a repair authorization number. Should the unit require return 
for service, it should be properly packed to avoid damage. Any damage resulting from improper packaging shall be the responsibility 
of the user.

Summary of Contents for I-4001-HCS

Page 1: ...Labnet Microtube Shaking Incubator Instruction Manual Catalog Numbers I 4001 HCS I 4002 HCS ...

Page 2: ...uipment Overview 4 7 0 Operation Guide 5 7 1 Before turning on the power 5 7 2 Setting up the of temperature vibration time and vibration speed 5 7 3 Pulse vibration 6 7 4 Correct temperature discrepancy 6 7 5 Buzzer set up 7 8 0 Troubleshooting and Repair 8 9 0 Spare Parts and Accessories 9 10 0 Limited Warranty 10 11 0 Equipment Disposal 10 ...

Page 3: ...kely to be damaged the rated safety level to be reduced or the rated operation conditions can be affected Labnet shall not be in any way responsible for the consequences resulting from an operator not observing the following requirements Grounding AC power s grounding should be reliable to safeguard against an electric shock The 3 pin plug supplied with the Labnet Microtube Shaking Incubator s pow...

Page 4: ...ating conditions that cause a high temperature environment will result in degraded performance or failure of the Labnet Microtube Shaking Incubator Additionally the device should be protected against any kind of heat sources such as sunlight ovens or central heating equipment If the Labnet Microtube Shaking Incubator will sit idle for extended periods of time it is recommended to disconnect the po...

Page 5: ... Temperature control range 0 C to 105 C Temperature range Room temperature 14 C to 100 C Timing range 1 min to 99 hrs 59 min Cooling time From room temperature to room temperature 10 C 8 min From 100 C to room temperature 10 C 15 min Heating time 12 min from 20 C to 100 C Temperature accuracy 0 5 C Temperature uncertainty 0 5 C Heating rate About 6 C min from 20 C to 100 C Block temperature unifor...

Page 6: ...ng Constant temperature Temperature decreasing Timer Vibration speed Set temperature Set vibration time Set vibration speed Actual temperature Actual vibration speed Actual vibration time Block Case main body Display panel and keyboard Sticker Power connector fuse and power switch ...

Page 7: ... to be abnormal when the power is turned on shut down the equipment immediately and contact the supplier The digits of the TEMP display may be reduced or increased by pressing Temp or Temp respectively The digits of the TIME display may be reduced or increased by pressing Time or Time respectively Press the above mentioned buttons for more than 3 seconds to set the exact value If the temperature s...

Page 8: ... TIME TEMP TIME SPEED TEMP TIME SPEED TEMP TIME SPEED TEMP TEMP TIME TIME TEMP TEMP TIME TIME TEMP TIME SPEED TEMP TIME SPEED TEMP ME SPEED TEMP TEMP TIME TIME TEMP TIME TEMP TIME TEMP TEMP TIME TIME TEMP TIME SPEED TEMP TIME SPEED TEMP TIME SPEED TEMP TEMP TIME TIME TEMP TIME SPEED TEMP TIME SPEED TEMP TEMP TIME TIME TEMP TIME SPEED TEMP TIME SPEED TEMP Temp Time TEMP TIME TIME TEMP TEMP TIME TIM...

Page 9: ...wing the same instruction when the temperature of the instrument increases to 100 C wait for 30 minutes input the calibrated value and press Short Prog to save it When the calibration of 40 C is done press Start Stop to exit At this time the calibration of 40 C is ineffective When the calibration is done once at both 40 C and 100 C check for the discrepancy between the two calibration points and t...

Page 10: ...ancy between the actual temperature and the displayed temperature Broken sensor or bad contact Contact the supplier or manufacturer The cooling of the module drastically slows down or the temperature cannot reach below room temperature Thermoelectric cooling module is broken Contact the supplier or manufacturer Fan is broken or does not work The module is neither able to heat nor cool Temperature ...

Page 11: ...250V 1 5A 5 x 20 mm 1 Accessories Cat No Description Quantity I 4000 A Block A holds 40 x 1 5 mL tubes 1 I 4000 B Block B holds 54 x 0 5 mL tubes 1 I 4000 C Block C holds 96 x 0 2 mL tubes 1 I 4000 D Block D holds 24 x 15 mm tubes 1 I 4000 E Block E water bath ID 115 x 73 x 38 mm not suggested to use when mixing function is on 1 I 4000 G Block G holds 26 x 0 5 mL and 24 x 1 5 mL tubes 1 I 4000 H B...

Page 12: ...esciences or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products received without proper authorization will be returned All items returned for service should be sent postage prepaid in the original packaging or other suitable carton padded to avoid damage Cor...

Page 13: ...listing of US medical devices regulatory classifications or specific information on claims visit www corning com resources www labnetlink com For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners 2022 Corning Incorporated All rights reserved 2 22 CLSLN AN 1001DOC REV2 Corning Incorporated Life Sciences www corning com lifes...

Reviews: