Labnet C2500-R User Manual Download Page 18

 

Prism R™ Refrigerated Microcentrifuge 

15 

 

To replace rotor, first make sure the motor shaft and rotor mounting 

hole are clean. Place the rotor on the motor shaft. Reinstall the rotor 

securing  screw  on  the  motor  shaft  by  turning  it  clockwise.  Hold  the 

rotor  with  one  hand  and  tighten  the  rotor  securing  screw,  using  the 

rotor wrench.

 

 

6.6. 

Replacing fuses 

Check fuse  when  it is  recommended in  the  Technical Specifications 

of  this  manual.  The  fuse  holder  is  located  in  the  power  inlet  on  the 

right-hand side of the unit. Disconnect the power cord from the power 

inlet.  Open  the  fuse  holder  drawer  by  inserting  a  small  screwdriver 

under the tab and prying it open. Remove the innermost (operative) 

fuse from its retaining tabs and replace the fuse if necessary. A spare 

fuse is located in the outer most chamber of the fuse drawer. Replace 

only with a fuse of the exact same value as the original.     

 

 

7.  Troubleshooting 

To  ensure  the  product’s  quality  and  performance,  the 

Prism  R

 

refrigerated microcentrifuge is 100% inspected by the manufacturer. 

Error messages will appear on the display if the 

Prism R

 refrigerated 

microcentrifuge fails to perform the attempted action properly.    In the 

case of Error (Err) messages or faults, please refer to the solutions in 

the following table to clear the error messages or faults.     

 

If the following solutions are not able to resolve the problem, please 

contact  Labnet International Inc.’  Service  Department to  arrange for 

Summary of Contents for C2500-R

Page 1: ...Lit M00502 July 2012 v2 C2500 R C2500 R 230V User Manual...

Page 2: ......

Page 3: ...w 8 5 Operation 10 5 1 Setting the Time TIME 10 5 2 Setting the Speed RPM 10 5 3 Setting the Centrifugal Force RCF 11 5 4 Setting the Temperature TEMP 11 5 5 Cooling 11 5 6 Quick Spin Operation 13 5 7...

Page 4: ...oid placing the machine in direct sunlight 5 Choose a flat stable surface capable of withstanding the weight of the machine 6 Install the machine in the room temperature 15 30 C relative humidity 20 8...

Page 5: ...least 30 cm of free space on all sides including the rear 15 Repair should be carried out by Labnet International authorized service personnel only 16 Use original spare parts and accessories only War...

Page 6: ...trol speed up to 13 500rpm 17 135xg Powerful refrigeration system maintain 4 C at max speed Includes unique easy access rotor 24 x 1 5ml Exceptionally quiet and compact Optional StripSpin adapter avai...

Page 7: ...ower source 100 120 50 60Hz countries while 3 15A fuse is for power source 200 230V 50 60Hz countries 3 2 Initial Operation Place the machine on a stable flat table Please keep the machine at least 30...

Page 8: ...tubes symmetrically The difference in the weight between the tubes should not exceed 0 1 gram The centrifuge tubes should be loaded in pairs opposite each other and each should contain roughly the equ...

Page 9: ...he START STOP knob When the pre set time expires the centrifuge will stop automatically To stop the centrifuge prior to the expiration of set time press the START STOP knob again When the time is set...

Page 10: ...l the lid is closed 3 6 Emergency Lid Release When the lid can t be opened by pressing the button users can open the lid manually First turn off the power of the centrifuge Second use a flat head scre...

Page 11: ...omponents and control panel see Figure 2 as well as the symbols and indicators on the LCD display see Figure 3 Figure 2 Prism R refrigerated microcentrifuge overview Figure 3 LCD display panel Air ven...

Page 12: ...g TEMP button To adjust temperature setting LID button Open the lid QUICK button Quick Spin centrifugation Audible Signals The audible signals are beneficial when you are familiarizing yourself with t...

Page 13: ...knob to increase or decrease the value Press button again to confirm the selection or press START STOP knob to store and run the selection The flashing of time value will stop and the time value will...

Page 14: ...value Press button again to confirm the selection or press START STOP knob to store and run the selection The flashing of RCF value will stop and the RCF value will be stored automatically if no butto...

Page 15: ...an be stopped by pressing the START STOP key The Pre Cooling will function only when the set temperature is lower than the ambient temperature Standby Cooling When the centrifuge is powered on the ref...

Page 16: ...ange the settings while the rotor is running Through a one time pressing of any function buttons the current value will switch into the pre set value mode The to be changed value flashes and can then...

Page 17: ...cabinet Maximum temperature 50 C 6 3 Cleaning the Rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately...

Page 18: ...er under the tab and prying it open Remove the innermost operative fuse from its retaining tabs and replace the fuse if necessary A spare fuse is located in the outer most chamber of the fuse drawer R...

Page 19: ...ame volume of liquid is in each tube Defective or improperly adjusted balance sensor Call for service Error rotor imbalance Rotor is stuck when a run starts Call for service Lid not closed correctly C...

Page 20: ...Centrifuge will not start Appendix A Technical Specifications Model C2500 R Max Speed 13 500 rpm Max RCF 17 135 g Acceleration Time 17 seconds Deceleration Time 18 seconds Temperature Range 10 C to 4...

Page 21: ...se and within this period all defective parts will be replaced at no charge by Labnet International Inc The warranty does not cover defects caused by excessive wear and tear or damage due to shipping...

Page 22: ...ice Name Prism R Refrigerated Microcentrifuge Device Models Numbers C2500 R C2500 R 230V Relevant EC Directives Low Voltage Directive 2006 95 EC EMC Directive 2004 108 EC RoHS 2002 95 EC WEEE 2002 96...

Page 23: ...AL OR CONSEQUENTIAL DAMAGE COMMERCIAL LOSS OR ANY OTHER DAMAGES RESULTING FROM THE USE OF THIS PRODUCT Some states do not allow limitation on the length of implied warranties or the exclusion or limit...

Page 24: ......