background image

4

3. Operation

ATTENTION:
Never operate the centrifuge  if the rotor shows any signs of

          corrosion or mechanical damage.

Never operate with strongly corrosive materials.

3.1

Closing the lid
After the rotor has been properly loaded, close the centrifuge

lid.

3.2

Lid  release
The "lid" button must be pushed to open the lid after a centrifuge run

has been completed.

3.3

Lid lock
The centrifuge can only be started with the lid closed.  Do not attempt to

open the lid until the rotor comes to a complete stop.

3.4

Starting the centrifuge
To start the centrifuge select the desired run time by turning the time knob
in a clockwise direction.  If performing a continuous run turn the time
knob counter clockwise into the "hold" position.
Once a run time is selected a given speed must be selected in order to
start the centrifuge.

The speed (rpm) can be selected between 1,000 and 4,000 rpm by turning

the knob to the 1, 2, 3 and 4 speed marks.  To select a speed, turn the

speed knob to the 2 posistion to activate, then proceed by turning the

knob to the desired speed.

At the end of the preselected time, the centrifuge stops automatically.

To stop the centrifuge prior to the selected time, turn the timer knob to

the "OFF" position.

2.4

Centrifuge Installation

1. Make certain that the timer is set for the "off" position.  Before
operating the centrifuge, check that the power source corresponds to
that on the manufacturer's rating label, then connect the power cord
to a properly grounded outlet

ATTENTION:
Ensure that the timer is in the "OFF" position, before connecting
the power cord.

!

NOTE:  The clinical centrifuge is calibrated at 120 volts line voltage.  Operating

the centrifuge at a line voltage other than 120V can significantlly impact the speed

settings below 2,000 RPM.

Summary of Contents for C0100-A

Page 1: ...INSTRUCTION MANUAL FOR THE Z100 A CLINICAL CENTRIFUGE C0100 A January 2005...

Page 2: ...3 2 4 CentrifugeInstallation 4 3 Operation 4 3 1 Closing the lid 4 3 2 Lidrelease 4 3 3 Lid lock 4 3 4 Startingthecentrifuge 4 4 Proper use of the Rotor 4 1 Overloading the rotor 5 4 2 Balancing the r...

Page 3: ...without the prior permission of the manufacturer Exception tubes and bottles made of glass or plastic normally available in the laboratory and designed for centrifugation The centrifuge is to be used...

Page 4: ...blocks until it has been established that your unit is working properly The accessories supplied with the centrifuge should be kept with the instruction manual near the centrifuge s place of installa...

Page 5: ...ected in order to start the centrifuge The speed rpm can be selected between 1 000 and 4 000 rpm by turning the knob to the 1 2 3 and 4 speed marks To select a speed turn the speed knob to the 2 posis...

Page 6: ...may operate a 6 place rotor with 2 3 or 4 loaded tubes but it is important that the occupied tube shields are symetrically arranged see sketch A 4 2 When loading the rotor you should proceed accordin...

Page 7: ...cohol or other disinfecting agents Do not allow the alcohol or disinfectant to sit on the units lid or housing as it may affect or discolor the plastic 5 4 Cleaning the centrifuge after breakage of gl...

Page 8: ...Check and set the time or speed Possible reason Blown fuse Solution Check fuse and replace if necessary see 5 5 Should you have any questions about the centrifuge and its accessories or require servic...

Page 9: ...alues See section 1 3 for the correct radius of your rotor Note the use of tube adapters will change thisradius To determine the relative centrifugal force RCF value Thisvalueisthemultipliertothegravi...

Reviews: