 8.0  Cleaning and Maintenance

Be sure to disconnect the power cord before cleaning or decontaminating the unit.

No routine maintenance is required for the electrical or mechanical components of the unit. The incubator exterior, interior, and 
shaking platform should be wiped down periodically with a soft damp cloth with mild soap. 

NOTE:

 Do not use chlorine-based bleach 

or abrasives. Any spills in the incubator and/or on the shaking platforms should be cleaned up immediately. 

 9.0  Troubleshooting and Service

Always make a visual inspection of the incubator and control panel when troubleshooting. Look for loose or disconnected wires that 
may be the source of trouble.

In the event the incubator does not operate properly, check the following before calling for service:

 



Is electrical power reaching the unit?

  – Check if the unit is plugged into the power supply.

  – Check if the circuit breaker has tripped.

  – Check if the Main Circuit Breaker has tripped.

 



Unit will not heat?

  – Verify the Main Control is set at the desired set point.

  – Verify the High Limit is set higher than the Main Control.

 



Unit will not maintain temperature?

  – Verify the Main Control is set at the desired set point.

  – Verify the High Limit is set higher than the Main Control. 

  – Check if the circulating fan is running.

  – Check the ambient room temperature. Fluctuation in ambient room temperature will have an effect on stability of the  

 

      incubators temperature.

Should you have a question about the Labnet 211DS Shaking Incubator or require service for the unit, contact Corning Customer service 
at 800.492.1110 or 978.442.2200. Do not send a unit for service without first calling to obtain a repair authorization (RA) number. The 
unit should be properly packed to avoid damage. Any damage resulting from improper packaging shall be the responsibility of the user.

Summary of Contents for 211DS

Page 1: ...Labnet 211DS Digital Shaking Incubator Instruction Manual Catalog Numbers I 5211 DS I 5211 DS 220V...

Page 2: ...0 Installation 2 5 1 Power Source 2 5 2 Location 2 5 3 Lifting Handling 2 5 4 Leveling 2 5 5 Cleaning 2 6 0 Controls 3 7 0 Operation 3 7 1 SmartCheck Temperature Control System 3 7 2 Calibration 3 7 3...

Page 3: ...roughly before operating the incubator All operators should be given adequate training before using the 211DS Shaking Incubator NOTE This equipment must be used only for its intended application Any a...

Page 4: ...rloading or circuit failure 5 2 Location In selecting a location consider all conditions which might affect performance such as heat from radiators ovens autoclaves etc Avoid direct sun fast moving ai...

Page 5: ...de on the controller press either the Up or Down arrow pad one time The digital display will start to blink going from bright to dim While blinking the digital display is showing the set point To chan...

Page 6: ...reaching the unit Check if the unit is plugged into the power supply Check if the circuit breaker has tripped Check if the Main Circuit Breaker has tripped Unit will not heat Verify the Main Control...

Page 7: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Page 8: ...isting of US medical devices regulatory classifications or specific information on claims visit www corning com resources www labnetlink com For a listing of trademarks visit www corning com clstradem...

Reviews: