4
| LABNET 211DS SHAKING INCUBATOR
8.0 Cleaning and Maintenance
Be sure to disconnect the power cord before cleaning or decontaminating the unit.
No routine maintenance is required for the electrical or mechanical components of the unit. The incubator exterior, interior, and
shaking platform should be wiped down periodically with a soft damp cloth with mild soap.
NOTE:
Do not use chlorine-based bleach
or abrasives. Any spills in the incubator and/or on the shaking platforms should be cleaned up immediately.
9.0 Troubleshooting and Service
Always make a visual inspection of the incubator and control panel when troubleshooting. Look for loose or disconnected wires that
may be the source of trouble.
In the event the incubator does not operate properly, check the following before calling for service:
Is electrical power reaching the unit?
– Check if the unit is plugged into the power supply.
– Check if the circuit breaker has tripped.
– Check if the Main Circuit Breaker has tripped.
Unit will not heat?
– Verify the Main Control is set at the desired set point.
– Verify the High Limit is set higher than the Main Control.
Unit will not maintain temperature?
– Verify the Main Control is set at the desired set point.
– Verify the High Limit is set higher than the Main Control.
– Check if the circulating fan is running.
– Check the ambient room temperature. Fluctuation in ambient room temperature will have an effect on stability of the
incubators temperature.
Should you have a question about the Labnet 211DS Shaking Incubator or require service for the unit, contact Corning Customer service
at 800.492.1110 or 978.442.2200. Do not send a unit for service without first calling to obtain a repair authorization (RA) number. The
unit should be properly packed to avoid damage. Any damage resulting from improper packaging shall be the responsibility of the user.