Labnet 211DS Instruction Manual Download Page 5

  Instruction Manual        | 

         3

 6.0  Controls

2

3

4

5 6

1

1.  Power switch: 

The main power (OFF/ON) switch controls all power to the unit. It must be in the ON position before any systems are 

operational.

2.  Shaker speed knob: 

Allows for speed adjustments of the shaker. Speed range is 20 to 400 rpm.

3.  Shaker power button: 

Controls the power to the shaker motor. It must be in the I/ON position before the shaker can operate. 

4.  SmartCheck™ Temperature Control System: 

Consists of the digital display and Up/Down arrow pads for inputting set point 

temperatures and calibration.

5.  Heating Indicator: 

This pilot light is On when the Temperature Control System has activated the heating element to reach and 

maintain the set point.

6.  Safety Light: 

This pilot light comes on when the Safety controller is activated. Under normal operating conditions this pilot light 

should never be On.

7.  Safety: 

This control, on the rear of the unit, is equipped with an adjustment knob and a graduated dial marked “0 to 10”. 

Completely independent of the Temperature Control System, the Safety guards against any failure which would allow temperature 
to rise past the set point. If temperature rises to the safety set point, the Safety takes control of the heating element and allows 
continued use of the incubator until the problem can be resolved or service can be arranged.

 7.0  Operation

Check the power supply voltage against the unit marking plate. They must match.

Plug the power cord into the grounded electrical outlet. Push the power switch to the ON position, and turn the Safety to its 
maximum position (10), clockwise.

7.1 SmartCheck Temperature Control System

Enter desired set point temperature. To enter set point mode on the controller, press either the Up or Down arrow pad 
one time. The digital display will start to blink, going from bright to dim. While blinking, the digital display is showing the 
set point. To change the set point, use the Up and Down arrow pads. If the arrow pads are not pressed for five (5) seconds, 
the display will stop blinking and will read the temperature of the unit. Allow the incubator at least 24 hours to stabilize.

7.2 Calibration

The unit was calibrated at the factory at 37°C, however it is recommended that the unit be recalibrated once it is in its working 
environment and has been stable at set point for several hours. Place a calibrated reference thermometer inside the chamber where it 
can be easily viewed. Make certain it is not touching any shelving or chamber walls. Allow the temperature to stabilize again until the 
temperature remains constant for 60 minutes. Compare the reading on the reference thermometer with the digital display. If there 
is a difference, put the display into calibration mode by pressing on both the Up and Down arrow pads at the same time and holding 
them in for about five (5) seconds or until the two outside decimal points start to flash. When the decimal points are flashing, the 
display can be calibrated to match the reference thermometer by pressing on the Up or Down arrow pads until the display reads the 
correct value. Allow the incubator temperature to stabilize, and repeat if needed.

7.3 Set Safety Controller

As mentioned above, the Safety should be initially set to its maximum position, to allow the unit to stabilize. Once the incubator is stable 
at the desired set point, turn the Safety counterclockwise until the Safety Light turns on. Then, turn the Safety clockwise just until the 
Safety Light turns off. This will set the Safety Controller at approximately 1°C above the SmartCheck Temperature Control System.

Summary of Contents for 211DS

Page 1: ...Labnet 211DS Digital Shaking Incubator Instruction Manual Catalog Numbers I 5211 DS I 5211 DS 220V...

Page 2: ...0 Installation 2 5 1 Power Source 2 5 2 Location 2 5 3 Lifting Handling 2 5 4 Leveling 2 5 5 Cleaning 2 6 0 Controls 3 7 0 Operation 3 7 1 SmartCheck Temperature Control System 3 7 2 Calibration 3 7 3...

Page 3: ...roughly before operating the incubator All operators should be given adequate training before using the 211DS Shaking Incubator NOTE This equipment must be used only for its intended application Any a...

Page 4: ...rloading or circuit failure 5 2 Location In selecting a location consider all conditions which might affect performance such as heat from radiators ovens autoclaves etc Avoid direct sun fast moving ai...

Page 5: ...de on the controller press either the Up or Down arrow pad one time The digital display will start to blink going from bright to dim While blinking the digital display is showing the set point To chan...

Page 6: ...reaching the unit Check if the unit is plugged into the power supply Check if the circuit breaker has tripped Check if the Main Circuit Breaker has tripped Unit will not heat Verify the Main Control...

Page 7: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Page 8: ...isting of US medical devices regulatory classifications or specific information on claims visit www corning com resources www labnetlink com For a listing of trademarks visit www corning com clstradem...

Reviews: