Operation
CoaDATA 501 – Operators Manual 2.1
Page 18
Quick
The curve for the PT calibration curve can be entered by typing in the 100
%-value and the factor for the slope of the calibration curve.
< quick >
• Press
Enter
to confirm selection.
<- 100 %
Example!
= 11.6 s
normal clotting time
factor = 54
Please use the value for
factor
as provided with the reagent package insert.
If you need to calculate the
factor
by yourself please refer to chapter 5.5.
If complete data has been entered for <curve> and <quick > the
calibration curve type selected last is active for con-versions in the
ready-to-measure mode. This is also valid for INR and ratio under
2nd conversion.
• Press
Esc
to access the measuring mode or
Enter
to add or verify
additional parameters.
none
If < none > was selected no conversion will be performed.
2nd conversion
< 2.conv >
If a calibration curve has been entered in menu 1st conversion or if <none>
has been selected, the 2nd conversion can be used for INR or ratio.
INR
• Press
Enter
and the following display appears:
< INR >
• Press
Enter
to type in the ISI-value.
ISI=
1.05
ATTENTION: If <none> was selected under 1st conversion the 100 % sec
value, e.g. 100 % = 12,6 sec. will be requested.
• Enter the ISI-value provided on the reagent package insert.
• Press
Enter
to confirm the entry.
NOTE