Refer to the following if your RapidVap Vertex Evaporator fails to operate
properly. If the suggested corrective actions do not solve your problem,
contact Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Unit will not operate
Unit not connected to
electrical power
Connect unit to proper electrical
receptacle.
Circuit breaker blown
Correct electrical problem and reset
circuit breaker by pressing button.
Lid open
Close lid.
Sample odor in lab
Vent hose exhausting into lab
area
Redirect hose to fume hood.
Inoperable blower
Replace blower.
Solvent in bottom of
chamber
Gas pressure too high
Reduce pressure.
Poor recovery rates
Sample going to dryness
Program time to prevent dryness.
Sample splashing out of tube
Decrease Nitrogen pressure.
Heat setting is too high
Decrease heat.