Now that you have an understanding of how to work in the Open Access Station,
we will review the suggested maintenance schedule and the common service
operations necessary to maintain your PuriCare for peak performance.
Many of the service operations should be performed only by
trained and experienced certification technicians after the
station has been properly decontaminated. DO NOT attempt to
perform these operations if you are not properly trained. The
service operations that require qualified certifiers are preceded
by the wrench icon.
Routine Maintenance Schedule
Weekly
Using 70% ethanol, or a suitable disinfectant, surface disinfect the inside of
the station, and the work surface.
Using an appropriate optical plastic cleaner and cloth, clean the sash.
Operate the station blowers, noting the pressure readings in an operational
log.
Lift the work surface and inspect the prefilters.
!