Example of Protector Filtered Fume Hood Settings Sheet
Labconco
Fan set points are based on ASHRAE 110 testing results conducted at Labconco.
Fan Set point
Dflt 2850 rpm
Fan Set point
Min 1500 rpm Variable Volume, 2850 rpm Constant Volume
Semicon/Solvent sensor
Trig. 3500mv
based on your LP questionnaire data LP #GFH-XXXXXXXXXX
Delay:
15 min
Electro/Acid sensor
Trig. Factory set based on sensor ppm value of the gel sensor itself
Delay:
15 min
Sash
Inverted Polarity
Trig.
65%
Days on filter:
Set it to the max of
730
days. This is always set to the maximum
of two years on every installation.
Activated audible alarms
All alarms are active (audible) out of the box.
K,
Solvent alarm threshold permanently exceeded. Saturated filters modules
N,
Acid alarm, threshold permanently exceeded. Saturated filter modules
Q,
End of acid sensor lifetime alarm
V,
40°C temperature alarm exceeded in the enclosure
W,
Top priority alarm 60°C temperature alarm threshold exceeded in the
enclosure
This data is needed for initial commissioning and if a reconfiguration is needed on
the hood during any service work.
Summary of Contents for Protector Echo 180 Series
Page 161: ...Appendix H Chemical Guide Product Service 1 800 522 7658 156 Original instructions...
Page 162: ...Appendix H Chemical Guide Product Service 1 800 522 7658 157 Original instructions...
Page 163: ...Appendix H Chemical Guide Product Service 1 800 522 7658 158 Original instructions...
Page 164: ...Appendix H Chemical Guide Product Service 1 800 522 7658 159 Original instructions...
Page 165: ...Appendix H Chemical Guide Product Service 1 800 522 7658 160 Original instructions...
Page 166: ...Appendix H Chemical Guide Product Service 1 800 522 7658 161 Original instructions...
Page 167: ...Appendix H Chemical Guide Product Service 1 800 522 7658 162 Original instructions...
Page 168: ...Appendix H Chemical Guide Product Service 1 800 522 7658 163 Original instructions...
Page 169: ...Appendix H Chemical Guide Product Service 1 800 522 7658 164 Original instructions...
Page 170: ...Appendix H Chemical Guide Product Service 1 800 522 7658 165 Original instructions...
Page 171: ...Appendix H Chemical Guide Product Service 1 800 522 7658 166 Original instructions...
Page 172: ...Appendix H Chemical Guide Product Service 1 800 522 7658 167 Original instructions...
Page 173: ...Appendix H Chemical Guide Product Service 1 800 522 7658 168 Original instructions...
Page 174: ...Appendix H Chemical Guide Product Service 1 800 522 7658 169 Original instructions...
Page 175: ...Appendix H Chemical Guide Product Service 1 800 522 7658 170 Original instructions...