Appendix E: gGuard
Product Service 1-800-522-7658
139
Original instructions
Check the GFH IP information
With your Administration Card, in the
“hood status”
menu of the GFH concerned by this
problem, check the IP Address, GW, and Subnet Mask to make sure it equals what you
have in gGuard
xxx.xxx.xxx.xxx
and what is on your lab diagram.
If 0’s or not correct you must re-enter the IP information from your lab diagram to that
specific hoods network parameters.
To do this go to the
“Administration Menu”
press
“OK”
to select it.
Then arrow down to
“Network Parameters”
press
“OK”
to select it.
On
“Ethernet”
press
“OK”
to select it.
Tip here select
DHCP
“Yes”
then
“No”
to get the subnet mask to partially fill out
if you have a static IP address.
Use the arrow keys to enter the IP information from your lab diagram. Once you
are sure it is correct press
“OK”
to
“save and exit.”
Now go back to the PC wit
h gGuard and “double click” on the hood in question. Once the
window opens to that hood press “F8” to force an immediate download of the file from the
hood.
FTP Server is locked
If the IP information is correct but you still cannot communicate with the hood you must
reboot the controller to unlock the FTP server.
To do this:
1. Press the light and fan icon on the controller and wait 30 seconds for it to turn off.
2. On the left side of the controller with the problem, pull out the black fuse cartridge
just enough to see the first fuse and break the power connection.
3. Wait 30 seconds and then push it back in. You will see it say
“Booting”
then
come up. Then turn the fans on with the “user” card.
4.
After it is on, go to the computer with gGuard and “double click” on the hood in
question. Once that window opens press “F8” to force an immediate download of
the file from the hood.
Summary of Contents for Protector Echo 180 Series
Page 161: ...Appendix H Chemical Guide Product Service 1 800 522 7658 156 Original instructions...
Page 162: ...Appendix H Chemical Guide Product Service 1 800 522 7658 157 Original instructions...
Page 163: ...Appendix H Chemical Guide Product Service 1 800 522 7658 158 Original instructions...
Page 164: ...Appendix H Chemical Guide Product Service 1 800 522 7658 159 Original instructions...
Page 165: ...Appendix H Chemical Guide Product Service 1 800 522 7658 160 Original instructions...
Page 166: ...Appendix H Chemical Guide Product Service 1 800 522 7658 161 Original instructions...
Page 167: ...Appendix H Chemical Guide Product Service 1 800 522 7658 162 Original instructions...
Page 168: ...Appendix H Chemical Guide Product Service 1 800 522 7658 163 Original instructions...
Page 169: ...Appendix H Chemical Guide Product Service 1 800 522 7658 164 Original instructions...
Page 170: ...Appendix H Chemical Guide Product Service 1 800 522 7658 165 Original instructions...
Page 171: ...Appendix H Chemical Guide Product Service 1 800 522 7658 166 Original instructions...
Page 172: ...Appendix H Chemical Guide Product Service 1 800 522 7658 167 Original instructions...
Page 173: ...Appendix H Chemical Guide Product Service 1 800 522 7658 168 Original instructions...
Page 174: ...Appendix H Chemical Guide Product Service 1 800 522 7658 169 Original instructions...
Page 175: ...Appendix H Chemical Guide Product Service 1 800 522 7658 170 Original instructions...