Appendix G: Acid/Formaldehyde Sensor Locations
Fan boxes that have acid or formaldehyde sensors will be labeled with small ‘A’ or ‘F’
stickers in the bottom left corner of the fan face and the carton it shipped in. The
installment locations for the sensors will be as follows:
Figure G-1
*
*
*
*
*
Key:
S
= Solvent Sensor (standard on all Fan Modules)
A
= Acid Sensor (optional)
F
= Formaldehyde Sensor (optional)
T
= Temp Sensor (standard on all Base Frames)
H
= Relative Humidity Sensor (standard on all Base Frames)
Notes:
1)
3-foot GFH: This filtered hood supports a max of (2) chemical
sensors. Available configurations are: S only, S + A, or S + F
2)
4-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 2 and Molecode F on
column 1.
3)
5-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 3 and Molecode F on
column 2.
4)
6-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 3 and Molecode F on
column 2.
5)
8-foot GFH: If both Molecode A and Molecode F are recommended,
Molecode A would be installed on column 4 and Molecode F on
column 3.
*Molecode S cell replacement for maintenance every 5 years
Summary of Contents for Protector Airo
Page 78: ...78 Fans and Lights not working ...
Page 79: ...79 Vertical sash no longer operates smoothly ...
Page 80: ...80 Electrical duplex outlets no longer have power Fan operates but lights dim or not working ...
Page 81: ...81 Contaminates outside the filtered hood ...
Page 82: ...82 Lights operate but fans will not ...
Page 83: ...83 Smart Command Alarms ...
Page 84: ...84 If needed contact Labconco to troubleshoot further ...
Page 94: ...94 Wiring Diagram Figure C 1 ...