Chapter 6: Maintaining the Protector Filtered Hood
Product Service 1-800-522-7658
92
Original instructions
Fan Box Replacement
Overview
1.
Remove front panels to expose the technology.
2.
Identify your failed fan box, 1, 2, 3, 4 or 5.
3.
Write down all parameters under “Hood status.”
4.
Turn off hood.
5.
Remove failed fan box.
6.
Put in new fan box.
7.
Follow the Reconfiguration process.
Step 1 – Remove front panels
Using your ladder, remove the panel or panels so you can access the RFID filter
tags later in this procedure.
Step 2 – Identify your failed fan box
As you stand in front of the hood looking left to right, the leftmost stack of
technology is fan #1, to the right of that is fan #2, and to the right of that fan #3, 4,
or 5, based on the size of your hood (M1, M2, M3, M4, or M5).
The micro switches are set accordingly based on their position, 1,2,3,4 or 5. So
identify what number the failed fan box is and make note of the
micro
switch
settings.
Make sure that the
micro
switch setting on the new fan box is set to match the
failed one’s settings that you are removing.
Step 3 – Write down parameters under “Hood Status”
Using a maintenance card, place it in the card reader and press
“Ok”
on the
“Hood status”
menu to open it.
It may be necessary to press
“Ok”
again
to open the
“Hood Status”
menu again.
Then arrow up or down to record the following on paper:
DHCP: X
IP: XXX.XXX.XXX.XXX
MSK: XXX.XXX.XXX.XXX
GW: XXX.XXX.XXX.XXX
Summary of Contents for Protector Airo 184 Series
Page 161: ...Appendix H Chemical Guide Product Service 1 800 522 7658 156 Original instructions...
Page 162: ...Appendix H Chemical Guide Product Service 1 800 522 7658 157 Original instructions...
Page 163: ...Appendix H Chemical Guide Product Service 1 800 522 7658 158 Original instructions...
Page 164: ...Appendix H Chemical Guide Product Service 1 800 522 7658 159 Original instructions...
Page 165: ...Appendix H Chemical Guide Product Service 1 800 522 7658 160 Original instructions...
Page 166: ...Appendix H Chemical Guide Product Service 1 800 522 7658 161 Original instructions...
Page 167: ...Appendix H Chemical Guide Product Service 1 800 522 7658 162 Original instructions...
Page 168: ...Appendix H Chemical Guide Product Service 1 800 522 7658 163 Original instructions...
Page 169: ...Appendix H Chemical Guide Product Service 1 800 522 7658 164 Original instructions...
Page 170: ...Appendix H Chemical Guide Product Service 1 800 522 7658 165 Original instructions...
Page 171: ...Appendix H Chemical Guide Product Service 1 800 522 7658 166 Original instructions...
Page 172: ...Appendix H Chemical Guide Product Service 1 800 522 7658 167 Original instructions...
Page 173: ...Appendix H Chemical Guide Product Service 1 800 522 7658 168 Original instructions...
Page 174: ...Appendix H Chemical Guide Product Service 1 800 522 7658 169 Original instructions...
Page 175: ...Appendix H Chemical Guide Product Service 1 800 522 7658 170 Original instructions...