Chapter 3: Getting Started
Product Service 1-800-522-7658
13
Original instructions
Connecting the Electrical Supply Source
to the Protector Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the
hood identification plate for the proper electrical requirements of your specific
model.
WARNING
:
The building electrical supply system for
Protector Hoods should include overload protection. A switch
or circuit breaker should be in close proximity to the
equipment and within easy reach of the operator. The switch
or circuit breaker is to be marked as the disconnecting device
for the equipment. Consult the current version of NFPA 70
®
,
NEC
®
for proper installation.
AVERTISSEMENT : Le système d’alimentation électrique de la
Hotte Protecteur doit inclure la protection contre la surcharge. Un
commutateur ou disjoncteur doit être tout près de l'équipement et
à portée facile de l'opérateur. Le commutateur ou le disjoncteur
doit être marqué comme l'appareil débranchant pour
l'équipement. Consultez la version actuelle de la norme NFPA 70
®
,
NEC
®
pour une installation correcte.
The identification plate, model number, serial number, and electrical connection
boxes are accessible from the front of the fume hood by removing the front panel.
The Protector Hood is normally wired for 115 Volt, 50/60 Hz, 20 Amp or 230Volt,
50/60 Hz, 10 Amp electrical service. Check the I.D. plate behind the front panel
for voltage verification. The number of circuits varies depending on the model.
All of the electrical connections for the individual duplexes are terminated at the
single point internal junction box for hook-up by a qualified electrician. If needed,
the individual duplexes can be converted for instant attachment to a wall outlet by
a qualified electrician by ordering 115V harness 9582700 or 230V harness
9582701. The main power to the control panel, fans, and lights uses less than 3
amps and a main power cord is included for instant attachment to a wall outlet.
The single point internal junction box is used for the connection of the duplex
outlets. Refer to the wiring diagram for your Protector Hood in
Appendix C:
Protector Filtered Hood Specifications.
The fume hood is required to be grounded to the MAINS protective earthing ground
for safe operation. Using a ring terminal sized for a 10-24 machine screw, connect
the MAINS ground conductor to the grounding lug marked with the protective
earthing symbol,
. Only MAINS ground conductors should be connected to the
protective earthing ground lug, no other conductors should be connected to this
grounding lug. Using wire nuts, connect the MAINS supply conductors to the fume
hood supply wires. Ensure that the wires are connected as per the appropriate wire
color codes for the input voltage. For 115V Phase (Hot) is black and Neutral is
!
Summary of Contents for Protector Airo 184 Series
Page 161: ...Appendix H Chemical Guide Product Service 1 800 522 7658 156 Original instructions...
Page 162: ...Appendix H Chemical Guide Product Service 1 800 522 7658 157 Original instructions...
Page 163: ...Appendix H Chemical Guide Product Service 1 800 522 7658 158 Original instructions...
Page 164: ...Appendix H Chemical Guide Product Service 1 800 522 7658 159 Original instructions...
Page 165: ...Appendix H Chemical Guide Product Service 1 800 522 7658 160 Original instructions...
Page 166: ...Appendix H Chemical Guide Product Service 1 800 522 7658 161 Original instructions...
Page 167: ...Appendix H Chemical Guide Product Service 1 800 522 7658 162 Original instructions...
Page 168: ...Appendix H Chemical Guide Product Service 1 800 522 7658 163 Original instructions...
Page 169: ...Appendix H Chemical Guide Product Service 1 800 522 7658 164 Original instructions...
Page 170: ...Appendix H Chemical Guide Product Service 1 800 522 7658 165 Original instructions...
Page 171: ...Appendix H Chemical Guide Product Service 1 800 522 7658 166 Original instructions...
Page 172: ...Appendix H Chemical Guide Product Service 1 800 522 7658 167 Original instructions...
Page 173: ...Appendix H Chemical Guide Product Service 1 800 522 7658 168 Original instructions...
Page 174: ...Appendix H Chemical Guide Product Service 1 800 522 7658 169 Original instructions...
Page 175: ...Appendix H Chemical Guide Product Service 1 800 522 7658 170 Original instructions...