Chapter 4: Lyo-Works Operating System
Product Service 816-333-8811 or 1-800-522-7658
29
Original instructions
Calibrate
- Pressing this button will save the entered offset value.
Temperature sensor offset values can range from -5 to +5° C.
Vacuum Sensor offset values can range from -0.020 to +0.020
mbar.
8.
Display
– This screen contains settings for controlling various functions of the
touch screen display.
Brightness Slider
– Moving the slider to the right will increase the
brightness level, moving to the left will decrease the brightness level
of the display.
Touch Sounds
– With Touch Sounds turned ON, there will be an
audible sound each time the display is touched. Pressing the Touch
Sounds button will toggle between ON & OFF.
Screen Timeout Function
– If this function is turned ON, the screen
will go into low power mode (sleep) when no screen touches have
been detected for the allotted set time (Screen Timeout). Pressing the
Screen Timeout Function button will toggle between ON & OFF.
Screen Timeout
– This value can be set to 5, 10, 15, 30 min, 1, 2, or 3
hours.
9.
Network
– If the DHCP Enabled button is turned ON, and an Ethernet cable is
connecting the Freeze Dryer to an active network, then the IP Address, Subnet
Mask, and Default Gateway fields will be populated automatically. If the
DHCP Enabled button is turned OFF, the IP Address, Subnet Mask, and
CALIBRATION
Offset:
________
Cancel
Calibrate