Chapter 5: Maintaining Your Freeze Dry System
Product Service 816-333-8811 or 1-800-522-7658
31
Monthly:
1.
The rubber components on the freeze dry system may eventually
deteriorate and require replacement. The effective life of rubber parts
depends upon both their usage and the surrounding environment. Check
all rubber hoses and gaskets and replace any that show signs of
hardening, permanent set or deterioration.
2.
Using a soft cloth, sponge or chamois and a mild, non-abrasive soap or
detergent, clean the acrylic door.
3.
Using a soft cloth, sponge, or chamois and a mild, non-abrasive soap or
detergent, clean the exterior surfaces of the cabinet. Liquid spray
cleaners and polishes may be used on the exterior surfaces. Do not use
solvents to remove stains from the exterior surfaces as they may damage
the finish.
Annually:
1.
Every 12 months, or more often if the freeze dry system is operated in
a dusty environment, the refrigeration system condenser should be
cleaned. Using a vacuum cleaner with brush attachment, clean the
condenser to ensure proper airflow for peak performance.