Chapter 4: Lyo-Works Operating System
Product Service 816-333-8811 or 1-800-522-7658
28
Original instructions
If all the SMTP fields entered are valid, but the network connection is not
working (or Ethernet cable is not connected) the following message will
appear after pressing the Save button:
EMAIL SETTINGS ERROR –
Network connection is not active
Test Email
– After the Email Settings and Network Settings have
been correctly entered, pressing this button will send a test email to
the selected email address.
Select Alerts
– The type of alerts that will be sent to each
individual email address can be customized. Pressing the Select
Alerts button will bring up the following selection list. Alerts can
be selected or deselected by pressing the check box in front of each
alert. Scroll list up or down by swiping screen up or down. Alerts
for the Freeze Dryer will be at the top of the list. Alerts for attached
accessories (Tray Dryer or Heated Product Shelves) will appear at
the bottom of the list.
6.
Maintenance
– This screen contains data and settings pertaining to the
operation hours and maintenance period for the vacuum pump and refrigeration
system. It also contains the Line Voltage Offset values. These offset values
determine the acceptable voltage range for electrical power supplied to the
Freeze Dryer. If the supply line voltage is outside of the acceptable range, then
the Line Voltage Range alert will be displayed. The default range is +/- 10%
of the nominal rated voltage (115V or 230V depending on model). The LOW
Line Voltage Offset can be adjusted from 0 to -10V (115V models) or 0 to -
20V (230V models). The HIGH Line Voltage Offset can be adjusted from 0 to
+10V (115V models) or 0 to + 20V (230V models). Scroll up and down to
view the entire page.