44
Product Service: Domestic 1-800-522-7658, International 816-333-8811
Chapter 7: Troubleshooting
III. System Components and
Collection Chamber Isolation
1. Remove the drying chamber or manifold.
2. Leave the gasket on and turn a large freeze dry
flask upside down and cover the connection
port.
3. Start the freeze dryer and observe the vacuum
indication. If the vacuum indication is good,
then the problem is in the drying chamber or
manifold and you should return to
Section II,
Gaskets, Tubing, Connections, Sample Valves
on page 42.
Freeze Dry Flask
Gasket
Freeze Dryer
Isolating the freeze dry system from the accessory chamber/manifold