7
8
Should you wish to change the temperature scale from centigrade to Fahrenheit enter the
menu as before and use the arrow icons/buttons until you get to a screen displaying
C F
.
Touch the tick icon/button to enter this function as before.
Centigrade/Fahrenheit
H/L
P
When you access the menu the first menu option is to view the set point, see above.
The set point is one of the parameters that you can not change but you can view it. Your
Labcold
I
ntelli
C
old® pharmacy refrigerator leaves the factory with a set point of 5oC as advised
by the Green Book chapter 3 page 27 and to prevent this being altered accidently, the set
point can only be changed by a
Labcold
engineer. If you need the set point changed please
contact the
Labcold
service team on
01256 705 580
or
To view the set point, touch the tick icon/button when the
SEt
screen is displayed and the
controller will display the set point temperature.
Viewing the Set Point
H/L
P
°C
H/L
P
To return to the normal display, touch the mute
icon/button or wait 30 seconds.
Viewing Software Version
This is another function that you cannot change, but you may be asked about the software
version if you need to contact us about an issue.
If the controller has reverted to the normal display, hold the arrow icons/buttons as above,
if you are already in the menu, touch the up and down arrows as required until you get to a
screen with
SFt
on it. Touch the tick icon/button as before and it will display a number. Again touch the
mute icon/button or wait 30 seconds for the display to revert to normal.
H/L
P
H/L
P
H/L
P
The
C
will flash, touch one of the arrows until the display shows
F
. Hold the tick icon/button
to confirm your selection and the screen will change from
F
to the the normal temperature
screen. The temperature will now display in Fahrenheit.
The temperature data for the Micro SD card will always be saved using the centigrade scale.
°
°t
1
F
H/L
P
Alarm Time Out
The alarm time out function allows you to change the time interval between the alarm being
muted and the alarm sounding again.
Access the menu by holding the arrow icons/buttons simultaneously and use either one of
them to scroll through options until you get to a screen that shows
Ato
.
Touch the tick icon/button to enter the function. The display will now flash with the current
number of minutes there are between the alarm being muted and sounding again. The fac-
tory setting is 15 minutes as shown below.
H/L
P
Use the arrow icons/button to adjust the interval in 5 minute units up to a maximum of a 30
minute time out. When you have set it to the length of time you require, hold the tick icon/
button and the screen will revert to
Ato.
You can then either leave it alone or hold the mute
icon/button and the controller will then revert to the normal temperature display.
H/L