SPECIFICATIONS: 

 

 

 

Enzyme/Manual Defrost Freezer PLUS Series 

  

  

  

Inner  

  

Usable Int. Dim. Ext. Dim. 

  

  

  

  

Ship

Model Cu.Ft. 

Doors 

Doors/Drawers Shelves 

W" 

D" 

H" 

W" D" H" 

H.P 

Amps 

BTU/hr 

Ibs

Ibs 

MFP-0920 

1 swing 

5/2 

6 fixed 

16 

16  50  22 24 64*  1/6 

850 

136 176 

MFP-2020 

20 

1 swing 

5/0 

4 fixed 

27 

18  57  32 29 77*  1/4 

770 

315 355 

MFP-2030 

20 

1 swing 

5/0 

4 fixed 

27 

18  57  32 29 77*  1/4 

770 

315 355 

MFP-2920 

29 

1 swing 

5/0 

4 fixed 

32 

26  65  35 32 84*  1/3 

2174 

328 368 

MCFP-1520 

15 

1 swing 

N/A 

2 baskets 

42 

24  28  48 30 35*  1/4 

770 

191 231 

MCFP-2020 

20 

1 swing 

N/A 

2 baskets 

56 

24  28  62 30 35*  1/4 

770 

220 265 

 
RECEIVING 

 

A. LRP ships by carriers that we have used for many years and have selected for their good service, 

but accidents in shipping occur regardless of how much care is taken.  Therefore the following is 
very important: 

 

B.  Inspect unit at once for any shipping damage, including concealed damage. 

 

C.  You should check for any shipping damage 

immediately

, preferably before the carrier leaves 

your receiving dock, as nearly all shipments are FOB ship point. 

 

D.  If there is any damage, note it on the carrier's receiving documents and notify the office of the 

carrier immediately. 

 

E.  The responsibility of the shipper and the delivering carrier ends five days after goods are received.  

It becomes the responsibility of YOU, the customer, to unpack the unit and check for concealed 
damage caused during shipping.  Immediate attention should be given to this. 

 

F.  Retain all packing materials if damage is found. 

 

INSTALLATION AND STARTING INSTRUCTIONS 

 

LOCATION – This unit requires a minimum of 4 inches of air flow in back and 1 to 2 inches on the sides 

and top.  Leveling legs are required in front only.  Unit may have a slight tilt from front to back after 
legs are installed.  This is acceptable and is recommended on this unit. 
 

POWER SOURCE – The supply circuit to this cabinet must conform to all National and Local Electrical 

Codes.  Consult the cabinet Serial-Data plate for voltage, cycle, phase, and ampere requirements 
before making connection.   VOLTAGE SHOULD NOT VARY MORE THAN 5% FROM SERIAL 
PLATE RATINGS.  A separate circuit is recommended to prevent possible loss of product due to 
over-loading or failure of other equipment on the same circuit.  PROTECT THE CIRCUIT WITH A 20 
AMP DELAY-TYPE FUSE.   
 

STARTING – There are no compressor shipping bolts to loosen or valves to open.  All that is necessary 

after the freezer has been properly leveled is to plug the service cord into an electrical outlet.  Do not 
use an extension cord.  Be sure your freezer is properly grounded.  Use the 3-prong plug provided 
into a 3-prong grounded outlet.   (Only this method complies with national electrical codes, local 
codes and ordinances.)  Unless the above grounding method is followed, you are not protected 
against severe or lethal shock in the event of a short circuit of an electrical component or wiring of the 
freezer.  When starting this new appliance, allow the cabinet to operate a minimum of three hours or 
until it has started cycling normally before placing product in the cabinet.  The motor compressor may 
start and stop several times when the unit is first started or after defrosting, especially if the weather 
is very hot.  This is only normal functioning of the motor overload protector.  The motor compressor 
will cycle normally as soon as the excess heat has been removed.    

Summary of Contents for Freezer PLUS Series

Page 1: ...Make thorough damage notation on your delivery receipt and have driver acknowledge by signature and date Send a written request asking for an inspection report from carrier Include the name of carrie...

Page 2: ...ions 1 RECEIVING 1 INSTALLATION AND STARTING INSTRUCTIONS 1 OPERATING INSTRUCTIONS 2 COMPRESSOR CONDENSER 2 SERVICE 2 OPERATION MANUAL FOR LR5 TEMPGUARD REFRIGERATION CONTROL 3 Lab Research Products P...

Page 3: ...inches of air flow in back and 1 to 2 inches on the sides and top Leveling legs are required in front only Unit may have a slight tilt from front to back after legs are installed This is acceptable a...

Page 4: ...ressor is hermetically sealed and will require no attention The condenser is located between the inside and outside walls of the freezer and also requires no attention Its natural heat keeps the exter...

Page 5: ...3 Operation Manual for LR5 TempGuard Refrigeration Control...

Page 6: ...re within the limits set by the user The control settings are easily adjusted through the LR5 TempGuard front panel and the control remembers the coldest and the warmest temperatures achieved in opera...

Page 7: ...ed How the LR5 Control works The LR5 TempGuard control continuously monitors the temperature of the refrigerator freezer through its sensor assembly When the temperature warms to the point where it re...

Page 8: ...D and the beeper Control Limits The High Temp and Low Temp can be adjusted within the entire range as desired The range varies depending on the model of refrigerator or freezer that was purchased The...

Page 9: ...ture gradient throughout the inner chamber The uniformity of the temperature throughout the inner chamber can be influenced by many factors such as location of the sensor air flow patterns mass of the...

Page 10: ...n contact Pin 6 Remote alarm common Pin 7 Remote alarm normally closed contact AC Line Power wired into AC line of the system Door alarm The door alarm circuit on the LR5 detects an open or closed cir...

Page 11: ...mpressor B Sensor Connector The LR5 sensor assembly connects here C Battery NiMH battery provides alarm capability when power is lost D 7 pin connector Wiring for LR 5 Tempguard system into connector...

Page 12: ...y output for remote alarm 12A max load for 120 VAC 10A max load for 240 VAC Visual red LED optional Audible optional Battery backed optional Door Open too long optional Temperature Detection DS1821 Di...