8
Wiring the LR5 TempGuard:
The LR5 Tempguard is supplied with a green 7-contact plug to make wiring easy. A
plug legend is attached to the back plate of the control. It is summarized below:
Pin 1
AC Line (polarity does not matter)
Pin 2
AC Line (polarity does not matter)
Pin 3
Door Alarm (polarity does not matter)
Pin 4
Door Alarm (polarity does not matter)
Pin 5
Remote alarm (normally open contact)
Pin 6
Remote alarm (common)
Pin 7
Remote alarm (normally closed contact)
AC Line
- Power wired into AC line of the system.
Door alarm
- The door alarm circuit on the LR5 detects an open or closed circuit.
An open circuit indicates that a door has been opened while a closed circuit
indicates that all doors are closed. The circuit should be wired in series through
switches attached at each door.
Remote Alarm
-
A relay is provided on the LR5 to signal an error condition.
Continuity exists between pins 6 & 7 in a normal condition. An error condition
provides continuity between pins 5 & 6.