
1
3
What to Do
Possible Reason
Trouble
Power switch is turned off.
Power does not come to the outlet.
Connect to the power source.
Switch it on.
Check the breaker, etc.
Connection of the tubes is wrong.
The artery tube is worn out.
The artery tube is worn out.
Air is collected in the circulation pump.
Replace it with a new artery tube.
(see page
6
,7,8,
9,
1
1
)
Replace it with a new tube.
(see page
6
,7,8,
9,
1
1
)
None of above is the case.
Skin covers are not set properly.
Adjust it by turning the adjustment knob clockwise.
(see page
10
)
The artery tube and/or skin covers are
worn out.
Set the cover again following the instruction manual.
(see page
9
)
The pulse pressure is too strong.
Replace it/them by new one(s).
Syringe is worn or the needle is clogged.
The pulse pressure is too weak.
Replace them with new syringe and needle.
The pulse pressure is too strong (too tight).
Loosen the adjustment knob by turning it counter-clockwise.
Tube sockets are not properly locked.
Check the sockets and lock them.
(see page 6,7,8
,9
)
Quick check-up before calling the customer service
Use the table if you have problems using the system. Look in this section for a description of the problem
to find a possible solution.
The system cannot be
powered on.
(Circulation pump does not
run)
Circulation pump runs but
simulated blood does not
flow.
Flash-back of the simulated
blood is not seen.
The simulated blood heavily
leaks from around the
puncture unit.
Bubbles appear in the
simulated blood in the tubes.
Power plug is not properly connected to
the power source.
Reset the connection following the instruction
manual.
(see page
7
)
Air is collected in the circulation pump.
The pump is sucking the bubbles coming
from the discharge tube.
Stop the circulation pump and push in water by
syringe .
(see page 1
2)
Adjust it by turning the adjustment knob
counter-clockwise.
Make a little distance between the ends of suction
and discharge tubes.
Discharge the air by tilting the pump as shown by
the illustration.
Make the pulse pressure weak (turn the knob
counter-clockwise), and run the pump until the
bubbles come off. Then return the pulse pressure
to an appropriate strength.
Simulated blood leaks from
the simulator base or pump.
Trouble
shooting
The tip of the suction / discharge tube is
out of the simulated blood.
Put in the tips of the both tubes to the simulated
blood.
(see page
10
)
Air accumulation is occurred in the tube(s).
Turning the knob counter-clockwise fully , pulse
pressure comes weak. Run the pump in this condition
until air accumulation is ejected.
(see page
10
)