― 4 ―
¡
Applicable standards
IEC 61557
IEC 61010-1 Measurement CAT.III 600V Pollution degree2
(Location for use: altitude 2000m or less)
IEC 61010-031
IEC 60529
IP40
¡
Measuring range and accuracy
<Insulation resistance range>
3.Specification
Nominal voltage
Auto Range
Open circuit
voltage
Short circuit
current
Nominal test
current
1
st
effective
measurement
range
Center scale
range
Accuracy
2
nd
effective
measurement
range
Accuracy
Accuracy at 0
KEW3021
1000V
0〜4/40/400/2000MΩ
Nominal test voltage 0%〜+20%
Within 1.5mA
1MΩ
1mA 0%〜+20%
0.100〜
1000MΩ
50MΩ
0〜
0.099MΩ
1001〜
2000MΩ
200.1〜
2000MΩ
40.0〜
2000MΩ
20.01〜
200.0MΩ
0〜
0.099MΩ
0〜
0.099MΩ
0〜
0.099MΩ
±(2%rdg+6dgt)
±(5%rdg+6dgt)
Within 6dgt
50MΩ
5MΩ
5MΩ
0.100〜
200.0MΩ
0.100〜
40.0MΩ
0.100〜
20.0MΩ
0.5MΩ
0.25MΩ
0.125MΩ
500V
250V
125V
0〜4/40/200MΩ
Nominal voltage
Auto Range
Open circuit
voltage
Short circuit
current
Nominal test
current
1
st
effective
measurement
range
Center scale
range
Accuracy
2
nd
effective
measurement
range
Accuracy
Accuracy at 0
KEW3022
500V
Nominal test voltage 0%〜+20%
Within 1.5mA
0.5MΩ
1mA 0%〜+20%
0.100〜
200.0MΩ
50MΩ
0〜
0.099MΩ
1001〜
2000MΩ
200.1〜
2000MΩ
40.0〜
2000MΩ
20.01〜
200.0MΩ
0〜
0.099MΩ
0〜
0.099MΩ
0〜
0.099MΩ
±(2%rdg+6dgt)
±(5%rdg+6dgt)
Within 6dgt
50MΩ
5MΩ
5MΩ
0.100〜
200.0MΩ
0.100〜
40.0MΩ
0.100〜
20.0MΩ
0.25MΩ
0.1MΩ
0.05MΩ
50V
0〜4/40/400/2000MΩ
250V
100V
0〜4/40/200MΩ