73
AC Current Measurement [RMS]
A
KEW1061
Rms-value detection, Crest factor: <3
Range
Resolution
Accuracy
Voltage Drop Maximum Input
Current
10 to 20Hz 20Hz to 1kHz
500μA
0.01μA
1.5+20
1+20
<0.11 mV/μA
440mA Protected
by a 440mA/
1000V fuse.
5000μA
0.1μA
50mA
0.001mA
<4mV/mA
500mA*3
0.01mA
5A
0.0001A
<0.1V/A
10A Protected by a
10A/1000V fuse.
10A
0.001A
AC Current Measurement [RMS]
A
KEW1062
Rms-value detection,
Crest factor: <3
Range
Resolution
Accuracy
Voltage
Drop
Maximum Input
Current
10 to 20Hz 20Hz to 1kHz 1k to 5kHz
500μA
0.01μA
1+20
0.75 +20
1+30
<0.11
mV/μA
440mA Protected
by a 440mA/
1000V fuse.
5000μA
0.1μA
50mA
0.001mA
<4 mV/mA
500mA *3
0.01mA
5A
0.0001A
1.5+20
1+20
2+30
<0.1 V/A
10A Protected by
a 10A/1000V fuse.
10A
0.001A
Model KEW1061/1062
Accuracy At 5 to 100% of range, At 10 to 100% of range for 10A Range
*3: Maximum measurement current : 440mA at 500mA range
Response time: 1 sec max.