INSTRUCTION MANUAL
KEW MATE 2000A/2001A
DIGITAL MULTIMETER
WITH
AC/DC CLAMP SENSOR
KYORITSU ELECTRICAL INSTRUMENTS
WORKS, LTD.
1.SAFETY WARNINGS
This instrument has been designed and tested according to IEC Publication 61010:
Safety Requirements for Electronic Measuring Apparatus. This instruction manual
contains warnings and safety rules which must be observed by the user to ensure
safe operation of the instrument and to retain it in safe condition. Therefore, read
through these operating instructions before starting using the instrument.
WARNING
● Read through and understand
instructions contained in this manual before
starting using the instrument.
● Save and keep the manual handy to enable quick reference whenever necessary.
● Be sure to use the instrument only in its intended applications and to follow
measurement procedures described in the manual.
● Be sure to understand and follow all safety instructions contained in the manual.
The instrument is to be used only in its intended applications.
Understand and follow all the safety instructions contained in the manual.
Failure to follow the instructions may cause injury, instrument damage and/or
damage to equipment under test. Kyoritsu is by no means liable for any damage
resulting from the instrument in contradiction to this cautionary note.
Failure to follow the above instructions may cause injury, damage to the instrument
and/or damage to equipment under test.
The symbol indicated on the instrument means that the user must refer to related
parts of the manual for safe operation of the instrument. Be sure to carefully read the
instructions following each symbol in this manual.
DANGER
is reserved for conditions and actions that are likely to cause serious or fatal injury.
WARNING
is reserved for conditions and actions that can cause serious or fatal injury.
CAUTION
is reserved for conditions and actions that can cause minor injury or instrument damage.
Following symbols are used on the instrument and in the instruction manual. Attention
should be paid to each symbol to ensure your safety.
Refer to the instructions in the manual.
This symbol is marked where the user must refer to the instruction manual so as not
to cause personal injury or instrument damage.
Indicates an instrument with double or reinforced insulation.
Indicates that this instrument can clamp on bare conductors when measuring
a voltage corresponding to the applicable Measurement category, which is marked
next to this symbol.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates AC and DC.
DANGER
● Never make measurement on circuits with a maximum voltage difference of
600VAC/DC or greater between conductors (300VAC/DC or greater between a
conductor and ground).
● Do not attempt to make measurement in the presence of flammable gasses.
Otherwise, the use of the instrument may cause sparking, which leads to an
explosion.
● Never attempt to use the instrument if its surface or your hand is wet.
●Do not exceed the maximum allowable input of measuring ranges.
● Never open the battery compartment cover while making measurement.
● Never try to make measurement if any abnormal conditions, such as broken
Transformer jaws or case is noted.
● The instrument is to be used only in its intended applications or conditions.
Otherwise, safety functions equipped with the instrument doesn t work, and
instrument damage or serious personal injury may be caused.
are noted,
to the
Return
under test
cover for
or color
before
e making
use. When
after
abrasives
shes safety
, and called
electrical
t designed
e designed
pply.
outlet by a
tly to the
tlets.
the power
nel).
that comes
not require
category
3.SPECIFICATIONS
●Measuring Ranges and Accuracy (at 23℃±5℃, relative humidity75% or less)
※Electromagnetic RF field ≦ 1 V/m
compatibility ACV/DCV/OHMS/FREQUENCY total accuracy = specified accuracy
(IEC 61000-4-3) ACA/DCA total accuracy = specified a5dgt
RF transmitters such as mobile telephones may not be used in close
proximity.
●Safety Standard
IEC 61010-1
measurement CAT
Ⅲ
, 300V, pollution degree 2
measurement CAT
Ⅱ
, 600V, pollution degree 2
IEC 61010-031
IEC 61010-2-032, IEC 61010-2-033
IEC 61326-1 (EMC), EN 50581 (RoHS)
●Operating System
Dual integration
●Display
Liquid crystal display with maximum reading of 3399 as
well as units and annunciators
Bar graph with maximum points of 33
●Over Input Indication
"OL" on the LCD (Ω ranges only)
●Auto-ranging Operation
Shifts to the next higher range when bar graph increases
to 33 points
Shifts to the next lower range when bar graph decreases
to 3 points
●Sample Rate Numeric reading: about 400ms,
bar graph: about 20ms
MODEL
2000A
2001A
Range
60A
100A
AC Current A
DC Current A
AC Voltage V Input impedance: 10MΩ
DC Voltage V Input impedance: 10MΩ
Measuring Range
0-60.0A
0-100.0A
Accuracy
± 2.0%rdg ± 5dgt(50/60Hz)
± 2.0%rdg ± 5dgt(50/60Hz)
MODEL
2000A
2001A
Range
60A
100A
Measuring Range
0- ± 60.0A
0- ± 100.0A
Accuracy
± 2.0% rdg ± 5dgt
± 2.0% rdg ± 5dgt
Range
3.4V
34V
340V
600V
Measuring Range
0-600V
(Auto-ranging)
Accuracy
± 1.5%rdg ± 5dgt(50-400Hz)
Range
340mV
3.4V
34V
340V
600V
Measuring Range
0- ± 600V
(Auto-ranging)
Accuracy
± 1.5%rdg ± 4dgt
Resistance Ω/
Range
340 Ω
3.4k Ω
34k Ω
340k Ω
3.4M Ω
34M Ω
Measuring Range
0-33.99M Ω
(Auto-ranging)
Accuracy
± 1.0%rdg ± 3dgt
Buzzer beeps below 30 ± 10Ω
(Continuity buzzer works on 340Ω range
only)
± 5%rdg ± 5dgt
± 15%rdg ± 5dgt
Frequency Hz
Range
Current
Voltage
Measuring Range
0-3.399kHz
3.4kHz-10kHz
(Auto-ranging)
0-3.399kHz
3.4kHz-33.99kHz
34kHz-300kHz
(Auto-ranging)
Accuracy
± 0.1%rdg ± 1dgt
± 0.1%rdg ± 1dgt
●Location for use
Indoor use, Altitude up to 2000m
●Accuracy-insured
23℃±5℃, relative humidity 75% or less
Temperature and
(without condensation)
Humidity Ranges
●Operating Temperature
0-40℃, relative humidity 85% or less
and Humidity Range
(without condensation)
●Storage Temperature
-20-60℃, relative humidity 85% or less
and Humidity Range
(without condensation)
●Source
Two 1.5VDC R03 (UM-4) batteries
●Current Consumption
Approx. 10mA
●Power-save Function
Shifts to the power-save state about 10 minutes after the
last switch operation
(current consumption: approx. 10μA)
●Overload Protection
AC/DC current ranges: MODEL2000A AC/DC 72A for
10 seconds
AC/DC current ranges: MODEL2001A AC/DC
120
A for
10 seconds
AC/DC voltage ranges: AC/DC 720V for 10 seconds
Resistance ranges: AC/DC 720V for 10 seconds
Frequency ranges: AC/DC 720V for 10 seconds
●Withstand Voltage
AC3470V for 5 sec. between electrical circuit and
housing case
●Insulation Resistance
10MΩ or greater at 1000V between electrical circuit and
housing case
●Conductor Size
MODEL2000A Approx. 6mm diameter max.
MODEL2001A Approx. 10mm diameter max.
●Dimensions
MODEL2000A 128(L)×87(W)×24(D)mm
MODEL2001A 128(L)×92(W)×27(D)mm
●Weight
MODEL2000A Approx. 210g
MODEL2001A Approx. 220g
●Accessories
Two R03 (UM-4) batteries
Instruction Manual
4.INSTRUMENT LAYOUT
Data Hold Button
MODEL2001 MODEL2000
Range Hold Button
Digital Multimeter
Function Selector Switch
Holster
Test Lead
LCD Display
AC/DC Clamp Sensor
Barrier
Protective fingerguard
DCA
Zero Adjust
Knob
5.PREPARATIONS FOR MEASUREMENT
(1)Checking battery voltage
Set the Function Selector Switch to any position other than the OFF position.
If the marks on the display is clearly legible without symbol
"BATT"
showing, battery
voltage is OK. If the display blanks or
"BATT"
is indicated, replace the batteries
according to section
8: Battery Replacement
.
NOTE
When the instrument is left powered on, the auto-power-save function automatically
shut the power off; The display blanks even if the Function Selector Switch is set to
a position other than the OFF position in this state. To power on the instrument,
turn the Function Selector Switch or press the Data Hold Button. If the display still
blanks, the batteries are exhausted. Replace the batteries.
(2)Make sure that the Function Selector Switch is set to the appropriate range.
Also make sure that data hold function is not enabled. If inappropriate range is
selected, desired measurement cannot be made.
(3)Install Test Lead to the Holster on the side of body
It is possible to measure with seeing the LCD Display keep
Test Lead installing to the Holster.
AC/DC
Battery Voltage Warning
Auto Range
Data Hold
Unit
Bar Graph
Buzzer
Manual
Range
Wind the lead around the folder.
Test Leads
AC/DC Clamp Sensor
6.HOW TO MAKE MEASUREMENT
6-1 Current Measurement
DANGER
● In order to avoid possible shock hazard, never make measurement on circuits with a
maximum voltage difference of 600VAC/DC or greater between conductors
(300VAC/DC or greater between a conductor and ground).
● Do not make measurement with the test leads connected to the circuit under test.
Never make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the barrier during measurement.
CAUTION
● When handling the clamp sensor, exercise caution not to apply excessive shocks or
vibration to the sensor.
●
M
aximum measurable conductor size is MODEL2000A 6mm / MODEL2001A
10mm in diameter.
6-1-1 DC Current Measurement
(1)The Function Selector Swith to the " A" position.
("DC" and "AUTO" marks are shown on the top of the display.)
(2) Turn the 0(Zero) ADJ knob to set the reading of the multimeter to zero. (If this zero
adjustment is made incorrectly, measurement errors will result.)
(3) Adjust one of the conductors to the center of the clamp sensor's arrow.
( When the position of the conductor is not at the center of the arrow, the error
occurs.)
Measured value is shown on the display.
Note: When current flows from the upside to the underside of the instrument, the polarity
of the reading is positive (+). Otherwise, the polarity of the reading is negative (-).
6-1-2 AC Current Measurement
(1)Set the Function Selector Switch to " A."
("AC" and "AUTO" marks are shown on the top of the LCD.)
(2) Adjust one of the conductors to the center of the clamp sensor's arrow.
( When the position of the conductor is not at the center of the arrow, the error
occurs.)
Measured value is shown on the display.
Note: Unlike DC current measurement, zero adjustment is not necessary. There is not
polarity indication either.
6-2 Voltage Measurement
DANGER
● In order to avoid possible shock hazard, never make measurement on circuits with a
maximum voltage difference of 600VAC/DC or greater between conductors
(300VAC/DC or greater between a conductor and ground).
● Do not make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the protective fingerguard during measurement.
6-2-1 DC Voltage Measurement
(1)Set the Function Selector Switch to " V."
("DC" and "AUTO" marks are shown on the top of the LCD.)
(2) Connect the red test lead to the positive (+) side of the circuit under test and the
black test lead to the negative (-) side. Measured voltage value is shown on the
display.
When the connection is reversed, "-" is shown on the display.
Conductor
MODEL2000A MODEL2001A
Adjust a
conductor to
the center of
the arrow.
Correct
Wrong
that the
value is
, the buzzer
value.
with a
under test.
under
Measured
is shown on
input
input
leads to
7.OTHER FUNCTIONS
7-1 Auto-Power-Save Function
NOTE
A small amount of current is consumed even in the power-save state. Make sure
to set the Function Selector Switch to the OFF position when the instrument is not
used.
This function helps to avoid unwanted exhaustion of the batteries because of
leaving the instrument powered on and extend battery life. The instrument
automatically shifts to the power-save state about 10 minutes after the last
Function Selector Switch or other switch operation.
To return to the normal state: Turn the Function Selector Switch or press the
Data Hold Button twice to exit the power-save state and enable measurement
functions.
7-2 Data Hold Function
This is a function to freeze a measured value on the display. Press the Data
Hold Button once to hold the current reading. In this data hold state, the reading
is held even if input varies. "H" and " " marks are shown on the LCD instead
of "AUTO" mark.
To exit the data hold state, press the Data Hold Button again.
7-3 Range Hold Function
The instrument defaults to auto-ranging ("AUTO" is shown on the LCD).
Pressing the Range Hold Button enables manual selection among measurement
ranges(" " mark is shown on the LCD instead of "AUTO" mark)
Press the Range Hold Button to select a higher range.
To switch from manual range selection to auto-ranging, press down the Range
Hold Button for about one seconds, or turn the Function Selector Switch to
another position before setting it back to the current range.
8.BATTERY REPLACEMENT
WARNING
● In order to avoid possible shock hazard, always disconnect the test leads from
the circuit under test and set the Function Selector Switch to the OFF position
before trying to replace the batteries.
CAUTION
● Do not mix new and old batteries.
● Install batteries in the orientation as shown inside the battery compartment,
observing correct polarity.
When the battery voltage warning mark "BATT" is shown on the top left corner of the
LCD, replace the batteries. Note that the display blanks and "BATT" mark is not
shown if the batteries are completely exhausted.
(1) Set the Function Selector Switch to "OFF."
(2) Remove the instrument from the holster.
(3) Loosen the battery-compartment-cover-fixing screw on the lower back of the
instrument.
(4) Replace the batteries with two new R03 (UM-4) 1.5V batteries.
(5) Put the battery compartment cover back in place and tighten the screw.
Screw
Battery
Compartment
Cover
Batteries
92-2318A
6-18
DISTRIBUTOR
Kyoritsu reserves the rights to change specifications or
designs described in this manual without notice and without
obligations.
Test Lead Cap:Test leads can be used under the CAT.II and III environments by
attaching a Protective cap as illustrated below. Use of our Protective cap offers
different lengths suitable for the test environments.
Barrer
Protective fingerguard: It is a part providing protection against electrical shock and
ensuring the minimum required air and creepage distances.
Protective Cap
Exposed metal part
Uncapped condition for CAT.II environment
Capped condition for CAT.III environments
This instrument satisfies the marking requirement defined in the WEEE
Directive. This symbol indicates separate collection for electrical and
electronic equipment.
This marking means they shall be sorted out and collected as ordained in
DIRECTIVE.
This directive is valid only in the EU. When you remove batteries from
this product and dispose them, discard them in accordance with domestic
law concerning disposal. Take a right action on waste batteries, because
the collection system in the EU on waste batteries are regulated.
www.
.com
1.800.561.8187