4. SPECIFICATION
Measurement Specification
Continuity
Open Circuit Voltage (DC) Short Circuit Current
Range
Accuracy
Greater than 6 V
Greater than
200mA@2Ω
20/200Ω
Auto - Ranging
Up to 2Ω
± (3%rdg + 4dgt)
Over 2Ω
± (3%rdg + 3dgt)
Insulation Resistance
Function Open Circuit Voltage (DC) Rated Current
Range
Accuracy
500V
500V+20%-0%
1mA or greater
@
500kΩ
20/200MΩ
Auto-Ranging
± (3%rdg + 3dgt)
1000V
1000V+20%-0%
1mA or greater
@
1MΩ
Loop Impedance
Rated Voltage
(AC)
Nominal Test Current at 0Ω External Loop
Range
Accuracy
230V+10%-15%
50Hz
25A/10ms
20Ω
± (3%rdg + 8dgt)
15mA/350ms max.
2000Ω
@ KAMP10 Test lead
RCD
Function
Rated
Voltage
(AC)
Test Current
Test
Current
Duration
Accuracy
Test Current
Trip Time
x1/2
230V
+10%
-15%
50Hz
10/30/100/300/500mA 2000ms
-8% -2%
± (1%rdg
+3dgt)
x1
10/30/100/300/500mA 2000ms
+2% +8%
FAST
150mA
50ms
DC
10/30/100/300mA
2000ms
±10%
500mA
200ms
Auto
Ramp
Goes up by 10% from 20%
to 110% of IΔn. 300ms x 10
±4%
Uc
Rated Voltage (AC)
Test Current
Range
Accuracy
230V+10%-15% 50Hz
5mA at IΔn=10mA
100V
+5% +15%rdg
±8dgt
15mA at IΔn=30/100mA
150mA at IΔn=300/500mA
Voltage Measurement
Rated Voltage (AC)
Measuring Range (AC)
Accuracy
100-250V 50Hz
100-300V
3% rdg
@ Loop/RCD/Uc Range