— 8 —
To prevent wrong connection of test leads and to maintain safety, the dedicated
terminals used for continuity and insulation tests are automatically covered when
using the terminals for Loop impedance, RCD and Uc tests.
Typical Number of Tests (central tendency for supply voltage up to 8V at R6P)
Continuity Ranges
: Approx. 700 times min. at load 1Ω
Insulation Resistance Ranges
:
Approx. 1000 times min. at load 0.5MΩ(500V)
Approx. 800 times min. at load 1MΩ(1000V)
LOOP/RCD/Uc Ranges
: Operational lifetime:5h (In case of continuous duty)
Operating instrumetal uncertainty
●
Operating instrumetal uncertainty of Continuity (IEC 61557-4)/Insulation Resistance (IEC 61557-2)
Function
Range
Measuring range to keep
operating uncertainty
Maximum percentage
operating uncertainty
Continuity
20Ω
200Ω
0.20-19.99Ω
20.0-199.9Ω
±30%
lnsulation
500V
0.50 - 199.9MΩ
Resistance
1000V
1.00 - 199.9MΩ
The influencing variations used for calculating the operating instrumetal uncertainty are denoted as follows;
Temperature
: 0
℃
and 35
℃
Supply voltage : 8V to 13.8V
● Operating instrumetal uncertainty of Loop Impedance (IEC 61557-3)
Range
Measuring range to keep
operating uncertainty
Maximum percentage
operating uncertainty
20Ω
0.4-19.99Ω
±30%
2000Ω
100-1999Ω
The influencing variations used for calculating the operating instrumetal uncertainty are denoted as follows:
Temperature
: 0
℃
and 35
℃
Phase angle
: At a phase angle 0° to 18°
System frequency : 49.5Hz to 50.5Hz
System voltage : 230V+10%-15%
Supply voltage
: 8V to 13.8V
● Operating instrumetal uncertainty of RCD (IEC 61557-6)
Function
Operating uncertainty of trip current
×1/2
-10% - 0%
×1, FAST
0% -+10%
Auto Ramp
-10% -+10%
The influencing variations used for calculating the operating error are denoted as follows:
Temperature : 0
℃
and 35
℃
Earth electrode Resistance (shall not exceed below) :
IΔn (mA)
Earth electrode resistance (
Ω
max.)
UL50V
UL25V
10
2000
2000
30
600
600
100
200
200
300
130
65
500
80
40
System voltage : 230V+10%-15%
Supply voltage : 8V to 13.8V