17
|
38
Atom series - v160120
Kvant Lasers s.r.o.
EN
1. Detach the cartridge from the cover and place it into the conventional oven - do not use microwave!!!
2. Heat the oven to 120ºC (250ºF) and leave the cartridge inside for about 2-3 hours, until all the moisture is released.
3. Insert the cartridge back into its place and close the optical compartment cover as soon as the cartridge cools down.
That way, it will absorb the maximum amount of moisture from the inside of the optical section.
The drying out process is pretty simple:
If the granules don’t turn brown during this process, they must be replaced with new ones.
!