![KUST Elektronik PT5040 Operation Manual Download Page 29](http://html1.mh-extra.com/html/kust-elektronik/pt5040/pt5040_operation-manual_2012318029.webp)
DIFF
Position
(
Limi
t
(
PHASE DIF
F
)
PT5040 Operation Manual Chapter 6 Introduction to [SYSTEM]
Volt
Step
Volt Step ranges from 1% to 50%. The step voltage is calculated as following formula:
Step volt = Start Volt * Volt Step.
Move the cursor to this zone, following soft keys will be displayed.
They are coarse soft keys used to change Volt Step by a resolution of 10%.
They are fine soft keys used to change Volt Step by a resolution of 1%.
Comparator
The comparator function allows you to set all comparators to ON or OFF. Once the comparator is
turned off, the instrument cannot make any comparisons, but the test wave or the standard wave
is also displayed on the measurement display page. Move the cursor to this zone, following soft
keys will be displayed.
ON Turn on the comparator.
OFF Turn off the comparator
Position
(AREA SIZE, DIFF ZONE, PHASE DIFF)
Set the comparison ranges for AREA SIZE, DIFF ZONE and CORONA that range from 0 to
6500 (the value of the end position must be larger than that of the start position.). Use numeric
buttons to enter the required value and press ENTER for confirmation. Move the cursor to this
zone, following soft keys will be displayed.
They are coarse adjustment soft keys used to change the value of position by 100.
They are fine adjustment soft keys used to change the value of position by 10.
)
Set the zero-crossing position for PHASE DIFF that is to select a zero-crossing position for the
comparison of phase difference. The zero-crossing position ranges from 2 to 99. Use numeric
buttons to enter the required value and press ENTER for confirmation. Move the cursor to this
zone, following soft keys will be displayed.
They are fine adjustment soft keys used to change the zero-crossing position by 1 dot.
)(
AREA SIZE, DIFF ZONE, PHASE DIFF
Set the limit values for comparison methods of AREA SIZE, DIFF ZONE, PHASE DIFF which
range from 0.1% to 99.9% by a resolution of 0.1%. Use numeric buttons to enter the required
value and press ENTER for confirmation. Move the cursor to this zone, following soft keys will
5-7