Volt
Samp
PT5040 Operation Manual Chapter 6 Introduction to [SYSTEM]
Imp Volt
The impulse voltage can be adjusted from 100V to 3000V(or 5000V) by a resolution of 10V.
Move the cursor to the impulse voltage zone, following soft keys will be displayed.
They are coarse adjustment keys used to increase and decrease the voltage by a
resolution of 100V.
They are fine adjustment keys used to increase and decrease the voltage by a
resolution of 10V.
Note: User can directly input the desired voltage by numerical keys. In the state
of inputting numbers, soft keys are unavailable. When the number is input,
ENTER can be used to end inputting and the default unit is volt (V).
le
200MSa/s, 100MSa/s, 50MSa/s, 20MSa/s, 10MSa/s, 5MSa/s, 2MSa/s, 1MSa/s, 500kSa/s,
200kSa/s, 100kSa/s are optional for sample rate.
Move the cursor to the sample rate zone, following soft keys are available.
Be used to select the desired sample rate.
Test
Imp
The number of tested impulses can be set from 1 to 32. Move the cursor to this zone, the
following soft keys will be displayed.
They are fine adjustment soft keys used to change the number of impulses by 1.
Note: User can directly input the desired number by numerical keys. In the state
of inputting numbers, soft keys are unavailable. ENTER can be used to
terminate inputting.
Erase
Imp
The number of erase impulses can be set from 0 to 16. Move the cursor to this zone, following
soft keys will be displayed.
They are fine adjustment soft keys used to change the number of impulses by 1.
Note: User can directly input the desired number by numerical keys. In the state
of inputting numbers, soft keys are unavailable. ENTER can be used to
terminate inputting.
ADJ
For different load on each DUT, when a given impulse voltage is applied across them, the actual