11
The specific operation
is as follows: The
figure on the right
shows the schematic
diagram of DC mA .
1. Insert the black
meter pen into the
“COM” jack and
the red meter pen
into the “V/
Ω
/Hz” jack;
the default value of the
current is DC, trigger the
“SELECT”
key to switch between
AC & DC current.
2.Turn the range switch to the corresponding DC/AC m A
gear, trigger “ SELECT” to select AC m A and then put
the meter into the circuit to be measured. The measured
current value and the current polarity of the red test point
will also be displayed at the same time. on the screen.
NOTE :
1.The instrument series connection to the circuit pending
to test before the circuit should be the first power off.
2.If there is no concept of the measured current range in
advance, the range switch should be transferred to the
3.000
AC
µ
A
SELECT
HOLD
MAX/MIN
REL
RANGE
FUSED
FUSED
CAT III 600V
10secMAX
10A
MAX
600mA
750V DC
1000V DC
10A
COM
V
W
HzmA
m
A
out
AUTO POWER OFF
TRUE RMS
OFF
V
V
~
NCV
W
Hz
%
A
mA
~
~
~
m
A
out
!
SELECT
µA
~
The Maximum measured pressure drop :
600mV
Overload Protection :
600
m
A / 6000
m
A : 600mA/250V Speed Glass Fuse.
AC CURRENT (
m
A)
600
m
A
6000
m
A
0.1
m
A
1
m
A
±(1.2%rdg + 10dgts)
Range
Accuracy
Resolution
12
highest range, & then according to the display value to
the corresponding file; such as the screen display “0L”
that has exceeded the range, needed to turn the range
switch to the appropriate gear.
3. The maximum input current is 600mA (depending on
the location of the red test pen inserted), excessive
current will damage the mA file fuse.
4.When the test leads are plugged into the current input
terminals, do not connect the test leads to any circuit in
parallel or it will damages the fuses and the instrument.
5.After the completion of all the measurement operation,
you should first turn off the power & then disconnect the
test leads & the measured circuit connection, Especially
to high current measurement.
6. It is forbidden to input more than 36V DC, 25V AC
Voltage between the current jack & the “COM” jack.
The Maximum measured pressure drop :
600mV
Overload Protection :
60mA / 600mA : 600mA/250V Speed Glass Fuse.
STRAIGHT 10A CURRENT (DC/ACA)
6 A
10 A
0.01 A
0.01 A
±(2.0%rdg + 30dgts)
Range
Accuracy
Resolution
The specific operation is as follows: The figure on the right shows
AC and DC 6A/10A
1.Insert the black meter pen into the “COM” jack and the red
meter pen into the “V/
Ω
/Hz” jack; the default value is DC, trigge-
ring the “RANG” key to switch between automatic / manual
6A &10A currents: