background image

12

Nearly  every  electrical  engineer  has  a  hand  held 

digital  clamp  meter  (Tongtester).  We  sometimes  take 
them for granted, until we damage them or “burn them 
out”.  If  you  incorrectly  connect  your  clamp  meter  to  a 
circuit, or if you have the clamp meter on wrong setting, 
you damage the meter and possibly hurt yourself. You 
can also get into trouble if you try to measure the voltage 
across a charged capacitor.

Clamp meter users frequently burn their meters by 

trying to measure current the same way as they measure 
voltage.  Remember,  you  measure  voltage 

across

  a 

circuit, and current 

through

 a circuit. When you use the 

current  input,  your  clamp  meter  becomes  a  low 
impedance circuit element.

Even if you correctly insert your clamp meter in to the 

circuit,  you  can  still  damage  your  meter.  Don’t  try  to 
measure  current  in  excess  of  your  meter’s  capacity. 
Check the current capacity of  the Clamp  meter first.

If  you  are  measuring  current  in  industrial 

environment,  to  prevent  excess  current  from  flowing 
through your meter, always disconnect your test leads 
from the circuit under test whenever you change Clamp 
meter functions. Set your meter to the correct function, 
say current, and its highest range for the setting. If the 
reading  is  small,  change  the  range  to  the  next  lower 
range till the reading can be read with the best possible 
accuracy.  When  measuring  voltage,  connect  the  test 
leads  before  your  apply  power  to  your  circuit.  To  be 
safe,  start  by  setting  your  meter  to  its  highest  range 
first.

TAKE MEASUREMENT CAREFULLY AND YOU’LL 

SPARE YOUR METER AND YOURSELF, SOME PAIN

WARRANTY

Each  “KUSAM-MECO”  product  is  warranted  to  be 

free  from  defects  in  material  and  workmanship 

under normal use & service.  The warranty period is 

one year (12 months) and begins from the date of 

despatch  of  goods.    In  case  any  defect  occurs  in 

functioning  of  the  instrument,  under  proper  use, 

within  the  guarantee  period,  the  same  will  be 

rectified by us free of charges, provided the to and 

fro freight charges are borne by you.

This warranty extends only to the original buyer or 

end-user customer of a “KUSAM-MECO” authorized 

dealer.

This  warranty  does  not  apply  for  damaged  Ic’s, 

fuses,  disposable  batteries,  carrying  case,  test 

leads, or to any product which in “KUSAM-MECO’s” 

opinion,  has  been  misused,  altered,  neglected, 

contaminated or damaged by accident or abnormal 

conditions of operation or handling.

“KUSAM-MECO” authorized dealer shall extend this 

warranty on new and unused products to end-user 

customers  only  but  have  no  authority  to  extend  a 

greater or different warranty on behalf of “KUSAM-

MECO”.

“KUSAM-MECO’s” warranty obligation is limited, at 

option,  free  of  charge  repair,  or  replacement  of  a 

defective product which is returned to a “KUSAM-

MECO”  authorized  service  center  within  the 

warranty period.

Summary of Contents for KM 2720

Page 1: ...Accuracy Overload Protection 250V rms Range Resolution 1 kW 4 MW 40 MW 10 kW 1 0 rdg 2dgts 1 0 rdg 2dgts 1 0 rdg 2dgts 1 0 rdg 2dgts 1 5 rdg 2dgts Range DIODE TEST Description Display read approx For...

Page 2: ...Transistorised Electronic Analog Digital Insulation Resistance Testers Digital Sound Level Meter Sound Level Calibrator Digital contact Non contact Type Tachometer Digital Non contact infrared Thermo...

Page 3: ...lectrical com THIS WARRANTY IS BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNE...

Page 4: ...eading can be read with the best possible accuracy When measuring voltage connect the test leads before your apply power to your circuit To be safe start by setting your meter to its highest range fir...

Page 5: ...tance Measurement Auto Power Off Maintenance 9 Battery Replacement 10 Fuse Replacement 10 Test Certificate 11 Warranty 13 11 MUMBAI TEST CERTIFICATE AUTORANGING CLAMPMETER This Test Certificate guaran...

Page 6: ...ith high voltage above DC 1000V orAC 750V 1 3 Be cautions when working at DC60V or AC30V such voltage may cause a shock hazard 1 4 When making measurements keep your finger behind the guard s plant on...

Page 7: ...a precis product Please do not promiscuously change the circuit or regulation resistance And pay attention to following 1 Do not use the Resistance Diode Transistor or Current range to test Voltage 2...

Page 8: ...01A 0 1A 0 01A 0 001A 40A 4A 2 0 rdg 2d 2 0 rdg 2d 0 01A 0 001A 400 mV 4 V 40 V 400 V Range AC VOLTAGE Model KM 2720 KM 2725 AC Accuracy 1 8 rdg 5d 750 V 1 0 rdg 3d 1 5 rdg 3d Note Input Resistance 1...

Page 9: ...tivity Frequency 0 7V AC RMS Voltage 0 1V AC RMS 0 2 rdg 3 DUTY Range 0 1 to 99 9 Unspecified Note Overload Protection 250V RMS Range RESISTANCE Accuracy 1 0 rdg 2d Note Overload Protection 250V RMS 4...

Page 10: ...circuit or component when the test leads are plugged into the current terminals 5 MEASUREMENT 5 1DC AC Current Measurement WARNING 1 Set the rotary switch to AC or DC 40A 200A position Then press REL...

Reviews: