Position Sensitivity

8

7

6

5

Observe the current value and waveform on the instrument’s display.
The flexible current probe can be used to AC current up to 3000A.
The probes provide a 3V full scale AC output proportional to the current 
being measured with three selectable ranges.
Select the corresponding sensitivity (.. mV/A) on your ScopeMeter or
multimeter. If you are using a multimeter, select an appropriate AC voltage 
range.
Don’t select AC+DC range for measurement.

Always use appropriate equipment for personal protection. 
Do not use the flexible current probe to measure bare conductors 
unless wearing protective clothing and gloves suitable for high 
voltage work.

Replacing Battery

1. When battery light is on, the battery needs to

 

be replaced.

2. 

Remove the flexible current 

    probe from any measuring 
    object.

3. 

Switch the main dial to “OFF”.

4. 

Open the battery door with a 

    screwdriver.

5. 

Replace the old batteries 

    with two new 1.5V (AAA Size) 
    batteries.

6. 

Close the battery door and 

    fasten the screw.

2 each 
AAA/R03 
batteries

Specifications

Probe Cable Length

(

s

)

:

Probe

 

Cable

 

Diameter:

Cable Length (Clamp to Box):

Cable Length (Box to Output):

Dimension (Box):

Weight:

254mm (10”)/ 458mm (18”)

8.5mm (nominal)

1.8m (6 ft)

0.5m

120mm(L)x70mm(W)x26 mm(H); 4.7”

(

L)x2.8”(W)x1”(H)

Approx.325g (include battery)

GENERAL SPECIFICATION:

(

At 23

±

5

80 %RH,

 

conductor located at the center of the flexible loop)

°C

ELECTRICAL SPECIFICATION:

Measuring Ranges:

Output Sensitivity:

Accuracy (45-65Hz):

Bandwidth(-3dB):

Phase Error (45-65Hz):

Position Sensitivity:

AC: 30A/300A/3000A

AC: 100mV/10mV/1mV per Amp.

±

1% of full scale

10Hz to 10KHz

±

1

°

Refer to Figure 1.

Noise:

Power Supply:

Power Indicator:

Low Battery:

Battery Life:

Load Impedance:

0.03A/0.075A/0.5A

UM-4 or AAA 1.5V battery x2

Indicated by a green LED

Indicated by a red LED

Approx.120Hours(alkaline battery)

10KΩ minimum

Temperature Coefficients:

Add 0.1 x specified accuracy for each degree C above 28

 or below 18

°C

°C

18”

Distance from optimum

Error

Flexible Current Probe

1 in

± 

1.0%

2 in
3 in

± 

2.0%

± 

3.0%

10”

Distance from optimum

Error

Flexible Current Probe

1 in

± 

1.0%

2 in

± 

2.5%

Note:

Measurement uncertainty assumes centralized primary conductor at 
optimum position, no external electrical or magnetic field, and within 
operating temperature range.

18”

10”

Measuring Instruction

Figure 2. Connecting the flexible probe 
around the  conductor

If the current probe is used with a multimeter, the actual current value 
can be calculated from the displayed AC voltage value.

Example 1 :

Current Probe set to 1 mV/A (3000A range). 
Multimeter displays 2.000Vac rms.

Example 2 :

Current Probe set to 10 mV/A(300A range). 
Multimeter displays 2.000Vac rms.

Figure 3. Measurement Setup

 

To activate unit, move the rotary 
switch from “ Off ” position to the 
required measuring range. 
Connect the banana plugs to the 
measuring meter. Select the desired 
range.

 

If the value of current being 

measured is unknown, first select the 
highest range and then reduce 
accordingly.
Connect the flexible probe around 
the conductor (See Figure 2.)
Center the conductor perpendicularly 
inside the flexible probe area.

Example 3 :

Current Probe set to 100 mV/A(30A range). 
Multimeter displays 2.000Vac rms.

actual current =                          =                 =                 = 2000A ac rms

display value

sensitivity

2.000V

1mV/A

2000mV

1mV/A

actual current =                          =                 =                 = 200A ac rms

display value

sensitivity

2.000V

10mV/A

2000mV
10mV/A

actual current                          

display value

sensitivity

2.000V

100mV/A

2000mV

100mV/A

=                          = 

=                     =  20A ac rms

Figure 1.

G 17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA. 

Sales Direct :

 (022) 24156638     

Tel. :

 (022) 24124540, 24181649, 27750662, 27750292    . 

Fax :

 (022) 24149659

Email :

 [email protected]             

Website : 

www.kusamelectrical.com

®

®

AN ISO 9001:2008 COMPANY

Reviews: