CA 3000

AC FLEXIBLE CURRENT PROBE

Symbol Definition & Button Location

1

2

3

7

Flexible current probe
Probe coupling
Power on / Range switch

Battery compartment cove

Name of Parts & Position:

Safety Information

Read the operating instructions and follow all safety instructions before 
using this instrument.

Do not operate the tester if the body of meter or the current probe looks 
broken.

Check the main function dial and make sure it is at the correct position 
before each measurement.

Always de-energize the circuit under test before installing flexible probe 
measuring head. Always inspect the electronics unit, connecting cable, 
and flexible probe for damage before using this product.

Always connect the clamp unit to display device before installing the 
flexible measuring head.

Exercise extreme caution when measuring live system with voltage 
greater than 60V DC or 30V AC.

To avoid false readings, replace the battery as soon as possible, when 

“        ”  LED light on.

Always wear protective clothing and gloves if hazardous live parts are 
present in the installation where the measurement is carried out.

Never use the probe on a circuit with voltages higher than 600V CAT III.

CAT III equipment is designed to protect against transients in equipment 
in fixed equipment installations, such as distribution panels, feeders and 
short branch circuits, and lighting systems in large buildings.

4

5

6

Banana plugs for output
Green LED - power indicator 

Red LED - low battery indicator

Environmental Conditions:

Altitude up to 2000 meters.

Operating temperature/humidity:

 

0

°

C~50

°

C, <80% RH, non-condensing

Storage temperature/humidity:-10

°

C ~ 60

°

C, <70% RH, battery removed

Pollution Degree: 2

2

4

3

Read first, 

To ensure safe operating and service of this current clamp, follow these 
instructions: 

Explanation of Symbols:

Attention! Refer to operation instructions.

Do not apply around or remove from HAZARDOUS LIVE conductors

.

Product is protected by double insulation

.

Conforms to relevant European standards

Approvals:

EN61010 

1000V CAT III / 600V CAT IV (Probe)
1000V CAT II / 600V CAT III (Box)

Maintenance

WARNING & 

!

CAUTION

Before opening the battery door, disconnect flexible current probe and 
never use the meter if the battery door is open.

Always inspect the unit, connecting cable, and flexible probe for damage 
before use.

To avoid electric shock, keep the clean and free of surface contamination.

Make sure the flexible probe, connecting cable, and electronics enclosure 
are dry before further use.

To avoid contamination or static damage, do not touch the circuit board 
without proper static protection.

REMARK:

CLEANING:

Periodically wipe the case with a dry cloth and detergent. 
Do not use abrasives or solvents on this instrument.

If

 

the

 

meter is not going to be used for a long time, take out the battery 

and do not store the meter in high temperature or high humidity environment.
When

 

taking

 

current

 

measurement

keep

 

the

 

cable

 

at

 

the

 

center

 

of

 

the

 

clamp

 

to

 

get

 

more

 

accurate

 

test

 

result

.

Repairs or servicing not covered in this manual should be performed only by 
qualified person

.