4
OFF
3000A
~
V
~
Hz
Ù
°C/°F
mV
~
400A
~
D. HOLD
Light
> 2 Sec
Push Button
REL
SELECT
Hz /
DUTY
MAX/MIN
Figure 3. Push Button
1. REL
r
?
Pressing this button, the meter enters relative
measuring mode, “REL
p
” is displayed on the
LCD and the present reading becomes the
reference value and displayed on the
secondary display. Relative measurement has
two modes.
REL
p
=
measurement
value-reference value, the other is REL%=
(REL
p
/Reference value) x100%(press
SELECT to select REL
r
or REL % mode)..
3
METER DESCRIPTION
q
Terminals
Refer to table 1 for terminal function
Function
Common terminal for all measurement
Terminal
COM
V/
Ù/Hz
Volts, Ohm, Diode, Freq., Temp.
and Cap. Measurement and square
wave output terminal
Table 1. Terminals
COM
V
Ù
Hz
1000V
750V
~
MAX
AC True RMS
DC / AC CLAMP
Figure 1. terminals
ROTARY SWITCH
Position of
Rotary switch
Function
400A
~
3000A
~
V
~
mV
~
Hz
Ù
°C/°F
DC& AC Ampere current
DC&AC Ampere current
DC&AC Voltage
DC&AC millivolts
Resistance, Diode & continuity
Capacitance
Temperature
Power off
Duty cycle / Frequency
Table 2.Rotary switch functions
off