Series 979 ATV Transducer
33
Protect Set Point – PRO
The PRO command returns or sets the pressure value at which the Hot
Cathode will turn itself off to prevent sensor damage. The protect set point
cannot be disabled and is valid during degas. NOTE: If the control set point is
enabled, do not set the protect set point below 3x10
-3
Torr.
Values:
1.0E-6 to 5.0E-2 Torr (default: 1.0E-2 Torr)
Query:
@001PRO?;FF
Query Response:
@001ACK1.0E-2;FF
Command:
@001PRO!5.0E-3;FF
Command Response:
@001ACK5.0E-3;FF
Calibration Commands
Atmospheric Calibration – ATM
The ATM command sets full scale readout for the MicroPirani. Vent the
transducer to atmospheric pressure before performing atmospheric
calibration.
For best results, leave the MicroPirani at the calibration pressure
for at least 20 minutes before using the ATM command.
Values:
Pressure value in scientific notation
Command:
@001ATM!7.60E+2;FF
Command Response:
@001ACK7.60E+2;FF
Vacuum Calibration – VAC
The VAC command zeroes the MicroPirani readout. Evacuate the transducer
to a pressure below 8x10
-6
Torr before performing vacuum calibration.
NOTE: The MicroPirani performs the vacuum calibration automatically when
the Hot Cathode pressure is below 4x10
-6
Torr.
For best results, leave the MicroPirani at the calibration pressure
for at least 20 minutes before using the VAC command.
Command:
@001VAC!;FF
Command Response:
@001ACKVAC;FF
Summary of Contents for 979 Series
Page 1: ...Operation and Maintenance Manual Series 979 Atmosphere to Vacuum Transducer ...
Page 2: ......
Page 3: ...Series 979 ATVTransducer Part 100014438 Series 979 Atmosphere to Vacuum Transducer ...
Page 8: ...Series 979 ATVTransducer ...
Page 48: ...48 Series 979 ATVTransducer Notes ...
Page 52: ...52 Series 979 ATVTransducer ...
Page 53: ...Series 979 ATVTransducer ...
Page 54: ...Series 979 ATVTransducer ...