Series 979 ATV Transducer
21
Pressure reading gas dependence: The MicroPirani is based on
measurement of thermal conductivity; therefore, the MicroPirani
readout depends on the gas type and concentration. The
MicroPirani is calibrated for Nitrogen gas, and will read a higher
pressure when exposed to atmospheric air. The Hot Cathode
sensor is based on measurement of gas ionization; therefore, the
Hot Cathode readout also depends on the gas type and
concentration.