Series 902 Piezo
Logarithmic 5
The following graph shows the correlation of logarithmic analog output to
pressure on a 1 to 5 volt scale. This is the output the user will see if the
value is
LOG5
. To receive analog output for
LOG5, the converter must have a 5 volt scale. The logarithmic output is
P (Torr) = 10
(V-2)
.
902 Logarithmic Analog Output
(1 - 5 Volt Scale)
0
1
2
3
4
5
0.1
1.0
10.0
100.0
1000.0
Pressure (Torr)
Voltage