KTS-Electronic GmbH & Co. KG
– GPA 500
18
5.1.2 Display after completion of data recording
Representation results from the two-dimensional image shown in 5.1.1
5.2 Display alternatives
5.2.1 Proportional or square display
With the button [prop.] the representation switches between proportional and
quadratic view. In the quadratic representation the display occurs with square
elements. Details partially are better recognized here. The relation between track
length to field width is not taken into account. The proportional representation is
displayed under consideration of the real track length and field width like it was setted
in [data recording].
Note
:
The actual measured length / width is only given through the setting [data
recording]. The correct input of the exact length and width is essential.
5.2.2 Data presentation - absolute or relative
There is the possibility to switch between absolute or relative data presentation. With
the absolute presentation the Z-axis is expanded to the full available measuring
range. With the relative presentation the value range is based on the smallest resp.
the largest measured value. This way a maximum detailed resolution is obtained
which also has effect on the color representation.