3.4.11 Zero calibration
The procedure for zero calibration is contained in the converter handbook. However, the
following information should be considered when installing the meter.
Zero calibration
1
Where the meter has been installed vertically, install shut-off valves either side of the meter to assist with zero cali-
bration.
2
If the process flow cannot be stopped, install a bypass section for zero calibration.