3
INSTALLATION
20
OPTIMASS 7000
www.krohne.com
10/2015 - 4001016303 - MA OPTIMASS 7000 R03 en
Use a regular pattern to tighten the bolts evenly
3.4.5 Maximum pipework forces (end loadings)
Mass flowmeters have a maximum level of force (negative or positive) that can be applied to the
ends of the meter. Refer to the table below for permitted forces.
Please refer to the table in the technical data section of this Handbook for the maximum end
loadings.
3.4.6 Pipework reducers
Always avoid extreme step changes in pipe size. Use pipework reducers, where there is a large
difference between pipework size and meter flanges.
MA_OPTIMASS_7000_en_151026_4001016303_R03.book Page 20 Monday, October 26, 2015 4:44 PM