SERVICE
7
27
OPTIFLUX 2070
www.krohne.com
04/2008 • 4000243601
•
Replace the cover.
•
Tighten the 4 bolts.
•
Enter the programming mode, hold the
V
and
Z
key for 5 seconds.
i
The display flashes.
•
Press the
Z
key to enter the menu.
•
Scroll through the available positions with the
V
key to position 74 and check the battery
capacity (important for battery life indication).
•
Go to position 75 and enter "1" (reset battery counter).
i
Battery capacity symbol at the display should be "full".
•
Hold the
V
key for 3 seconds to confirm the new value and go back to the measuring mode.
•
Replace the blue protection cap.
WARNING!
Make sure that the battery cable is not jammed by the cover.