Start-up
5.1 Configuration with Toolbox
The meter can be configured using the supplied software and adaptor, which allows connection
to a personal computer (PC) or a laptop.
Connection with Toolbox
•
Connect the meter to a personal computer or laptop using the converter.
•
Launch Toolbox but DO NOT try to connect to the meter.
•
Connect the meter to the power supply.
•
Select either Connection > Connect or Connection > Auto Connect to connect Toolbox to the
meter. This MUST be done within 10 seconds of energising the meter.
INFORMATION!
When the meter is in configuration mode, the pulse output is not available.
1
Personal computer or laptop
2
Toolbox software CD
3
Connection plug
4
Mass flowmeter
5
RS458 to USB converter