background image

Magnetic resonance multiphase flowmeter 

M-PHASE 5000

M-PHASE 5000

M-PHASE 5000

M-PHASE 5000

Supplementary Instructions

Supplementary Instructions

Supplementary Instructions

Supplementary Instructions

© KROHNE 04/2016 - 4004673401 - AD M-PHASE5000 Ex R01 en

Summary of Contents for M-PHASE 5000

Page 1: ...hase flowmeter M PHASE 5000 M PHASE 5000 M PHASE 5000 M PHASE 5000 Supplementary Instructions Supplementary Instructions Supplementary Instructions Supplementary Instructions KROHNE 04 2016 4004673401...

Page 2: ...ning the documentation 3 1 1 4 Warnings and symbols used 4 1 1 5 Manufacturer 4 1 2 Safety instructions for the operator 5 1 3 Approvals 6 1 4 General 6 1 5 Marking label 8 2 Temperature limits 10 2 1...

Page 3: ...es of such changes 1 1 2 Product liability and warranty The operator shall bear responsibility for the suitability of the device for the specific purpose The manufacturer accepts no liability for the...

Page 4: ...rking with electricity DANGER This warning refers to the immediate danger of burns caused by heat or hot surfaces DANGER This warning refers to the immediate danger when using this device in a hazardo...

Page 5: ...cal equipotential bonding system of the potentially explosive area Components attached or installed junction boxes bushings Ex type cable glands connectors shall be of a technical standard that compli...

Page 6: ...0 2012 EN 60079 1 2014 EN 60079 7 2007 EN 60079 18 2015 EN 60079 28 2015 1 4 General The electronic boxes Ex d contains the following screw type Cylinder head screw Standard ISO4762 Size M8x16 Materi...

Page 7: ...inals for Power and Signal connections all in type of protection Ex e For electrical data of these terminals refer to section 3 2 The Ethernet is terminated on a connector at the outside of the field...

Page 8: ...nameplate with device details as shown in the following two figures Figure 1 1 Location and details of nameplate 1 Information about manufacturer and product 2 Information about CE and explosion prot...

Page 9: ...9 M PHASE 5000 www krohne com 04 2016 4004673401 AD M PHASE5000 Ex R01 en Figure 1 2 Location and details of nameplate 1 Information about manufacturer 2 Information about explosion protection 3 Warni...

Page 10: ...k The instrument is not heated up by any additional heat radiation direct solar radiation heat from adjacent plant parts so causing it to operate above the permissible ambient temperature range Insula...

Page 11: ...ed cables must be used The shield must be connected to the EMC cable gland The mains power cable has three wires There are different colors used to identify the wire as in table below INFORMATION The...

Page 12: ...the equipotential bonding system of the hazardous area See IEC 60079 14 clause 6 3 For this purpose the internal or external connection facility PE clamp can be used Figure 3 1 Signal cables and power...

Page 13: ...orted This means that KROHNE can only service an instrument if it is accompanied by the following certificate confirming that the instrument is safe to handle If the instrument has been operated with...

Page 14: ...ress Department Name Tel No Fax No The enclosed device Type KROHNE Order No or Series No has been operated with the following liquid Because this liquid is water hazardous radioactive toxic caustic fl...

Page 15: ...ax No Manufacturer s order no or serial no The device has been operated with the following medium KROHNE Order No or Series No has been operated with the following liquid Because this liquid is water...

Page 16: ...rvices Head Office KROHNE Messtechnik GmbH Ludwig Krohne Str 5 47058 Duisburg Germany Tel 49 203 301 0 Fax 49 203 301 10389 info krohne com KROHNE 04 2016 4004673401 AD M PHASE5000 Ex R01 en Subject t...

Reviews: